iStent insertion orientation and impact on trabecular meshwork motion resolved by optical coherence tomography imaging.
Gong, Zhaoyu; Johnstone, Murray A; Wang, Ruikang K.
Afiliación
  • Gong Z; University of Washington, Department of Bioengineering, Seattle, Washington, United States.
  • Johnstone MA; University of Washington, Department of Ophthalmology, Seattle, Washington, United States.
  • Wang RK; University of Washington, Department of Bioengineering, Seattle, Washington, United States.
J Biomed Opt ; 29(7): 076008, 2024 Jul.
Article en En | MEDLINE | ID: mdl-39070082
ABSTRACT

Significance:

The iStent is a popular device designed for glaucoma treatment, functioning by creating an artificial fluid pathway in the trabecular meshwork (TM) to drain aqueous humor. The assessment of iStent implantation surgery is clinically important. However, current tools offer limited information.

Aim:

We aim to develop innovative assessment strategies for iStent implantation using optical coherence tomography (OCT) to evaluate the position and orientation of the iStent and its biomechanical impact on outflow system dynamics.

Approach:

We examined four iStents in the two eyes of a glaucoma patient. Three-dimensional (3D) OCT structural imaging was conducted for each iStent, and a semi-automated algorithm was developed for iStent segmentation and visualization, allowing precise measurement of position and orientation. In addition, phase-sensitive OCT (PhS-OCT) imaging was introduced to measure the biomechanical impact of the iStent on the outflow system quantified by cumulative displacement (CDisp) of pulse-dependent trabecular TM motion.

Results:

The 3D structural image processed by our algorithm definitively resolved the position and orientation of the iStent in the anterior segment, revealing substantial variations in relevant parameters. PhS-OCT imaging demonstrated significantly higher CDisp in the regions between two iStents compared to locations distant from the iStents in both OD ( p = 0.0075 ) and OS ( p = 0.0437 ).

Conclusions:

Our proposed structural imaging technique improved the characterization of the iStent's placement. The imaging results revealed inherent challenges in achieving precise control of iStent insertion. Furthermore, PhS-OCT imaging unveiled potential biomechanical alterations induced by the iStent. This unique methodology shows potential as a valuable clinical tool for evaluating iStent implantation.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Malla Trabecular / Algoritmos / Tomografía de Coherencia Óptica Límite: Humans Idioma: En Revista: J Biomed Opt Asunto de la revista: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Malla Trabecular / Algoritmos / Tomografía de Coherencia Óptica Límite: Humans Idioma: En Revista: J Biomed Opt Asunto de la revista: ENGENHARIA BIOMEDICA / OFTALMOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos