Safety and feasibility of continuous monitoring of coronary flow velocity in acute myocardial infarction by Doppler guide wire.
Am Heart J
; 132(3): 567-71, 1996 Sep.
Article
en En
| MEDLINE
| ID: mdl-8800026
We investigated the safety and limitations of the Doppler guide wire for continuous monitoring of coronary flow velocity outside the catheterization laboratory in 17 patients with acute myocardial infarction. After direct angioplasty, patients were taken to the coronary care unit with the Doppler guide wire positioned within the artery. Flow velocity was continuously monitored. Heparin was titrated to an active clotting time > 200 seconds. Clinical outcome and angiographic analyses were evaluated. Flow velocity monitoring was conducted with an 88.2% success rate and lasted for 16 +/- 5 hours. Monitoring failed in 2 of the 4 right coronary artery cases. Small amounts of thrombus were seen to adhere to the Doppler guide wire at the end of monitoring. No complications were related to the procedure. No deterioration of angiographic findings was observed. This preliminary study confirmed the safety of the Doppler guide wire for continuous monitoring of coronary flow in patients with acute myocardial infarction.