Your browser doesn't support javascript.
loading
Interference of hemolysis on glycated albumin determined by ketoamine oxidase method and its correction / 临床检验杂志
Article en Zh | WPRIM | ID: wpr-514024
Biblioteca responsable: WPRO
ABSTRACT
Objective To investigate the effects of hemolysis on glycated albumin (GA) determined by ketoamine oxidase method and its correction.Methods GA concentration and hemolytic parameter(optical density,A) in non-hemolytic serum and different degree hemolytic serum samples were measured.The impact of hemolysis on GA and the relationship between hemolysis and GA were analysed.A formula was developed to correct the interference of hemolysis on GA measurement using regressive Multiple analysis.Results Compared with non-hemolytic serum,hemolysis resulted in the significantly decreased concentrations of GA detected by ketoamine oxidase method(P < 0.01),which were significantly associated with the degree of hemolysis (R2 =0.943 4).Y and Z represented GA concentration of non-hemolytic serum and different degree hemolytic serum,while X represented optical density of hemolytic parameter.Formulas for GA measurement were presented:Y =2.468X + Z-0.015 73,GA concentrations of hemolytic samples can be reverted to the values without statistical difference from the GA concentration in corresponding non-hemolytic samples.The bias of corrected GA was less than 10%.Conclusion Our results indicate that the level of GA measured through ketoamine oxidase method is negatively affected by hemolysis.The formula of mathematical correction of GA results in hemolytic samples should be suitable for the requirements of clinical laboratory.
Palabras clave
Texto completo: 1 Base de datos: WPRIM Idioma: Zh Revista: Chinese Journal of Clinical Laboratory Science Año: 2017 Tipo del documento: Article
Texto completo: 1 Base de datos: WPRIM Idioma: Zh Revista: Chinese Journal of Clinical Laboratory Science Año: 2017 Tipo del documento: Article