Estimulación eléctrica permanente en región septal en pacientes con insuficiencia cardíaca secundaria a estimulación en ápex de ventrículo derecho / Permanent septal pacing in patients suffering secondary heart failure compared to right ventricular apical pacing
Chávez-González, Elibet; Barja, Luis D; Ortega, Daniel F; Pérez-Rodríguez, Alexis.
Afiliação
  • Chávez-González, Elibet; Universidad de Ciencias Médicas. Facultad de Medicina. Departamento de Electrofisiología y Estimulación Cardiaca. Santa Clara. CU
  • Barja, Luis D; Hospital Universitario Austral. Servicio de Electrofisiología y Marcapasos. Buenos Aires. AR
  • Ortega, Daniel F; Hospital Universitario Austral. Servicio de Electrofisiología y Marcapasos. Buenos Aires. AR
  • Pérez-Rodríguez, Alexis; Universidad Médica Dr. Serafín Ruiz de Zárate Ruiz. Facultad de Medicina. Santa Clara. CU
Arch. cardiol. Méx ; Arch. cardiol. Méx;88(5): 474-482, dic. 2018. graf
Article em Es | LILACS | ID: biblio-1142159
Biblioteca responsável: BR1.1
RESUMEN
Resumen

Introducción:

Independientemente de la modalidad de estimulación (VVI o DDD), el mayor porcentaje de estimulación ventricular derecha deteriora la función contráctil izquierda. Por ello se han estudiado diferentes sitios de estimulación.

Objetivo:

Describir las diferencias electrocardiográficas y ecocardiográficas entre la estimulación en la región apical y la región septal del ventrículo derecho (VD)0.

Métodos:

Se estudió a 24 pacientes, 2 mujeres y 22 hombres, con fracción de eyección deprimida (≤ 35%) secundaria a estimulación en punta de VD. Se realizó electrocardiograma durante estimulación en ápex de VD y en seguimiento de estimulación septal. También ecocardiografía en ambos momentos.

Resultados:

La fracción de eyección se incrementó desde 31 ± 3.1% hasta 45 ± 12% (p = 0.0041) con estimulación septal; esta mostró mejor grado de sincronía mecánica.

Conclusiones:

La estimulación septal puede constituir un sitio de elección en pacientes con disfunción sistólica de ventrículo izquierdo secundaria a estimulación en ápex de VD, los cuales presenten QRS estrecho en su conducción intrínseca, pues en la muestra estudiada produce una mejoría en la sincronía electromecánica, demostrándose incrementos significativos en la fracción de eyección.
ABSTRACT
Abstract

Introduction:

Regardless of the type of electrical stimulation (VVI or DDD) the highest percentage of right ventricular apical pacing can cause left ventricular failure. For this reason, studies have been performed in different sites on right ventricle pacing.

Objective:

To describe differences between electrocardiography and echocardiography variables during right ventricular apical pacing and septal pacing.

Methods:

A total of 24 patients were studied, 2 women and 22 men, with heart failure due to conventional pacing on right ventricular (ejection fraction ≤ 35%). An electrocardiogram as well as an echocardiogram, was performed during right ventricular apical pacing and when patients were paced on septal area.

Results:

The ejection fraction increased from 31 ± 3.1% to 45 ± 12% (P=.0041) on septal pacing, showing higher degree of mechanic synchronisation.

Conclusions:

Pacing on septal area could be a good site for those patients that suffer heart failure due to right ventricular apical pacing. These must show narrow QRS on their intrinsic electrocardiographic conduction. This kind of pacing can produce an improvement in electro- mechanical synchronisation, as well as show an increased left ventricular ejection fraction.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Estimulação Cardíaca Artificial / Função Ventricular Esquerda / Disfunção Ventricular Esquerda / Insuficiência Cardíaca Tipo de estudo: Etiology_studies Limite: Adult / Aged / Female / Humans / Male Idioma: Es Revista: Arch. cardiol. Méx Assunto da revista: CARDIOLOGIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Argentina / Cuba País de publicação: México

Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Estimulação Cardíaca Artificial / Função Ventricular Esquerda / Disfunção Ventricular Esquerda / Insuficiência Cardíaca Tipo de estudo: Etiology_studies Limite: Adult / Aged / Female / Humans / Male Idioma: Es Revista: Arch. cardiol. Méx Assunto da revista: CARDIOLOGIA Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Argentina / Cuba País de publicação: México