Your browser doesn't support javascript.
loading
Respirometriá por impedancia eléctrica transtorácica bipolar / [Respirometry by bipolar electric transthoracic impendance]
Clavin, O E; Valentinuzzi, M E.
Afiliação
  • Clavin, O E; s.af
  • Valentinuzzi, M E; s.af
Acta physiol. latinoam ; 28(6): 243-53, 1978.
Artigo em Espanhol | LILACS-Express | BINACIS | ID: biblio-1158606
Biblioteca responsável: AR1.1
Localização: [{"text": "AR1.1"}]
ABSTRACT
In a population of ten young normal male subjects (average height of 174.2 cm +/- 3.8, average weight of 70.2 kg +/- 6.16), which was subjectively classified as mainly ectomorph (slightly biased towards the mesomorph type) a good correlation was found between the transthoracic impedance changes due to normal respiratory movements and tidal volume (average r of 0.963 +/- 0.034) and a mean coefficient b1 of 3.64 ohms/liter +/- 1.82. For equal volume, inspiratory and expiratory movements, on the average, produced impedance changes with no significant difference (p less than 0.55). Attempts to quantitatively correlate the coefficient b1 (ohms/liter) with the anthropometric parameters did not produce any consistent result.
Buscar no Google
Coleções: Bases de dados nacionais / Argentina Base de dados: BINACIS Idioma: Espanhol Revista: Acta physiol. latinoam Ano de publicação: 1978 Tipo de documento: Artigo
Buscar no Google
Coleções: Bases de dados nacionais / Argentina Base de dados: BINACIS Idioma: Espanhol Revista: Acta physiol. latinoam Ano de publicação: 1978 Tipo de documento: Artigo
...