Avaliação in vitro das microdeformações em implantes curtos, com conexão protética cone morse e diferentes desenhos de roscas, convencionais e platôs / In vitro evaluation of micro-strain implants in short prosthetic connection with different designs and morse taper thread, conventional and plateaus
São José dos Campos; s.n; 2014. 88 p. ilus, tab, graf.
Tese em Português | BBO - Odontologia | ID: biblio-867574
Biblioteca responsável: BR243.1
Localização: BR243.1; D74, S85a
RESUMO
O objetivo desse estudo in vitro foi avaliar por meio da extensometria, a distribuição de microdeformações ao redor e em dois níveis verticais de implantes curtos cone morse com roscas convencionais e implantes com platôs. Em cada bloco de poliuretano (n = 10), foram instalados dois implantes com cada tipo de desenho. Para o implante de roscas convencionais, o pilar protético sólido, foi parafusado com torque de 20 N.cm e para o implante com platôs, o pilar protético reto foi assentado no interior do implante com auxílio de um martelete. Quatro extensômetros foram colados tangenciando cada implante, posicionados diametralmente opostos. Também foi colado um extensômetro no terço médio e outro no terço apical. Carregamentos estáticos axiais de 30 kg durante 10 segundos foram realizados sobre os pilares protéticos, com auxílio do dispositivo de aplicação de carga (DAC). Os dados obtidos na extensometria foram submetidos à análise estatística ANOVA e Sidak’s (5%). Os resultados mostraram que o tipo de implante (p = 0,00029 < 0,05), a região (horizontal e vertical) (p = 0,0001 < 0,05), assim como o efeito interação (p = 0,02044 < 0,05) foram estatisticamente significantes. Conclui-se que as microdeformações são maiores para o implante com platôs e que as microdeformações horizontais são maiores comparadas as verticais de ambos os implantes
ABSTRACT
The aim of this in vitro study was to evaluate by means of strain gauge analysis, the stress distribution around and in two vertical levels of short morse taper threaded implants with conventional screws and Bicon plateaus implants through gauge test. In each polyurethane block (n=10), two implants with each type of design were installed. For the conventional screws implant, the solid abutment was screwed with a torque of 20 N.cm and for the plateau implant,the abument was seated right inside the implant with the aid of a hammer. Four strain gauges were glued tangent to each implant, positioned diametrically opposed. Two strain gauges were also adhered to the middle third and the other one in the apical third. Axial static loads of 30 kg were performed on the abutments during 10 seconds,with the aid of a load device (DAC). The data obtained in the gauge test were submitted to statistical analysis using ANOVA and Sidak´s tests (5%).The results showed that the implants(p=0.00029 < 0.05), the region(horizontal and vertical) (p= 0.0001 <0.05), as well as the interaction effect(p=0.02044<0.05) were statistically significant. It follows that the micro-strains are greatest for the implant with plateaus and horizontal micro-strain are larger compared to both the vertical implants
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais Base de dados: BBO - Odontologia Assunto principal: Fenômenos Biomecânicos / Implantação Dentária / Implantação Dentária Endóssea Limite: Humanos Idioma: Português Ano de publicação: 2014 Tipo de documento: Tese
Texto completo: Disponível Coleções: Bases de dados nacionais Base de dados: BBO - Odontologia Assunto principal: Fenômenos Biomecânicos / Implantação Dentária / Implantação Dentária Endóssea Limite: Humanos Idioma: Português Ano de publicação: 2014 Tipo de documento: Tese
...