Your browser doesn't support javascript.
loading
Primera etapa del Programa de Estandarización de Creatinina en la Provincia del Chaco: relevamiento de tecnologías y procedimientos de los laboratorios clínicos / First stage of the Creatinine Standardization Program in the Province of Chaco: clinical laboratories technologies and procedures survey
Velasco, Gustavo; Bianchi, María Eugenia; Cusumano, Ana María; Susan Tauguinas, Ana Maria; Mazziotta, Daniel.
Afiliação
  • Velasco, Gustavo; Hospital Dr. Julio C. Perrando A. Argentina
  • Bianchi, María Eugenia; Universidad Nacional del Nordeste. Argentina
  • Cusumano, Ana María; Centro de Educación Médica e Investigaciones Clínicas. Argentina
  • Susan Tauguinas, Ana Maria; Fundación Renal del Nordeste Argentino. Argentina
  • Mazziotta, Daniel; Fundación Bioquímica Argentina. Argentina
Rev. nefrol. diálisis transpl ; 33(1): 25-33, mar. 2013. tab, graf
Artigo em Espanhol | BINACIS | ID: bin-130078
Localização: AR444.1
RESUMEN

Introducción:

El Ministerio de Salud Pública, la Fundación Renal del Nordeste Argentino, el Colegio Bioquímico, todas instituciones de la Provincia del Chaco, y la Fundación Bioquímica Argentina emprendieron el desarrollo de un Programa de Estandarización de la Creatinina Plasmática (PECrP). El objetivo es presentar los resultados del relevamiento realizado en los laboratorios clínicos (LC).

Métodos:

Se realizo una encuesta al 100% (n= 39) de los LC que participan en el programa, siendo del sector publico 22 (56%) los cuales fueron seleccionados por complejidad, tasa mensual de determinaciones de Creatinina y cercanía a poblaciones en desventaja y a los 17 (44%) LC privados que aceptaron participar.

Resultados:

Los resultados fueron 1) Tecnología automática 43,7%, semiautomática 38,4%, manual 17,9%; 2)

Metodología:

en la modalidad punto final 12,9% y cinética 79,2%, no respondieron 3 (7,69%) laboratorios; 3) Calibradores 46,3% utilizan patrones en matriz sérica, 17.9% patrones acuosos y no se obtuvieron datos del 35.8%; 4) Los sistemas homogéneos se constataron en el 12,9% de LC, mientras que en el 84,6% se utilizaron sistemas heterogéneos; 5) Temperatura el 41% hacían la reacción a 25 ºC, mientras que el 48,7% a 37 ºC; 6) Controles de calidad Internos (CCI) y Controles de calidad externos (PEEC) solo el 38,4.% utiliza ambos, el15,4% ninguno y el 25,6% utiliza solo una de las dos herramientas (CCI o PEEC), resultando un 41% de los laboratorios participantes con controles parciales o ausentes.

Conclusiones:

Teniendo en cuenta la heterogeneidad observada en los instrumentos y en las condiciones de la determinación se definieron nuevas líneas de trabajo para continuar con el programa de estandarización.(AU)
ABSTRACT

Introducción:

The Ministry of Public Health, La Fundación Renal del Nordeste Argentino (The Renal Foundation of the Argentinean Northeast), El Colegio Bioquímico (The BiochemicalSchool), all the institutions of the Province of Chaco, and La Fundación Bioquímica Argentina (The Argentinean Biochemical Foundation) beganthe development of a Plasmatic Creatinine Standardization Program (PCrSP). The objective is to present the results of the survey conducted in the Clinical Laboratories (CL).

Methods:

A survey was performed in 100% (n= 39) of the CL participating in the program; 22 (56%) of them belong to the public sector and were selected by complexity, monthly rate of creatinine determinations, and proximity to vulnerable populations, and 17 (44%) of them are private CL that agreed to participate.

Results:

The results were as follows 1) Technology automatic (44.8%), semi-automatic (37.9%), manual (17.2%); 2)

Method:

5 (13.9%) used endpoint and 31 (86.1%) used kinetic method; 3 laboratories did not provide a response; 3) Calibrators 18 (36.7%) used serum matrix, and 7 (17.9%) aqueous patterns, and no data was obtained from 14 (35.9%) labs; 4) Homogeneous systems were verified for 5 (13.2%) labs, and for 33 (86.8%) of them systems were heterogeneous; 5) Temperature 16 (43.2%) reacted at 25 ºC, and 19 (51.4%) at 37 ºC; 6) Internal Quality Controls (IQC) and External Quality Controls (EQC) 15 (48.3%) performed both, 6 (19.4%) performed neither, 10 (32.3%) performed either of them (IQC or EQC), where 51.7% of the participating laboratories had partial or absent controls.

Conclusions:

Considering the heterogeneity observed in both the tools and the determination conditions, new operating lines for standardization were defined.(AU)
Assuntos

Buscar no Google
Coleções: Bases de dados nacionais / Argentina Base de dados: BINACIS Assunto principal: Técnicas de Laboratório Clínico / Creatinina Tipo de estudo: Estudo diagnóstico / Guia de prática clínica Idioma: Espanhol Revista: Rev. nefrol. diálisis transpl Ano de publicação: 2013 Tipo de documento: Artigo Instituição/País de afiliação: Centro de Educación Médica e Investigaciones Clínicas/Argentina / Fundación Bioquímica Argentina/Argentina / Fundación Renal del Nordeste Argentino/Argentina / Hospital Dr. Julio C. Perrando A/Argentina / Universidad Nacional del Nordeste/Argentina
Buscar no Google
Coleções: Bases de dados nacionais / Argentina Base de dados: BINACIS Assunto principal: Técnicas de Laboratório Clínico / Creatinina Tipo de estudo: Estudo diagnóstico / Guia de prática clínica Idioma: Espanhol Revista: Rev. nefrol. diálisis transpl Ano de publicação: 2013 Tipo de documento: Artigo Instituição/País de afiliação: Centro de Educación Médica e Investigaciones Clínicas/Argentina / Fundación Bioquímica Argentina/Argentina / Fundación Renal del Nordeste Argentino/Argentina / Hospital Dr. Julio C. Perrando A/Argentina / Universidad Nacional del Nordeste/Argentina
...