Your browser doesn't support javascript.
loading
Desempeño analítico de los métodos de control de calidad de una solución concentrada para hemodiálisis / Analytical performance of the quality control methods in a concentrated solution for hemodialysis
Benavides Pérez, Juan Carlos; Rojas Romaguera, Rosa; Rodríguez Amado, Jesús Rafael; Lafourcade Prada, Ariadna.
Afiliação
  • Benavides Pérez, Juan Carlos; Laboratorio Farmacéutico Oriente. Santiago de Cuba. Cuba
  • Rojas Romaguera, Rosa; Laboratorio Farmacéutico Oriente. Santiago de Cuba. Cuba
  • Rodríguez Amado, Jesús Rafael; Laboratorio Farmacéutico Oriente. Santiago de Cuba. Cuba
  • Lafourcade Prada, Ariadna; Laboratorio Farmacéutico Oriente. Santiago de Cuba. Cuba
Rev. cuba. farm ; 47(4)oct.-dic. 2013.
Artigo em Espanhol | CUMED | ID: cum-58998
Biblioteca responsável: CU1.1
RESUMEN

Objetivo:

evaluar el desempeño de los métodos de cuantificación de los iones sodio, potasio, cloruro, calcio y magnesio en una solución concentrada para hemodiálisis con bicarbonato.

Métodos:

los cinco métodos aparecen informados en la farmacopea británica. Para la verificación del desempeño, se utilizaron los parámetros, linealidad, exactitud y precisión. Se empleó el método de placebo enriquecido utilizando niveles de 60 a 140por ciento de la concentración nominal de cada ion.

Resultados:

los cinco métodos resultaron lineales en el intervalo de concentraciones evaluadas, fueron precisos y exactos, con coeficientes de variación global menor que 1,75 por ciento, y una recuperación entre 100 ± 3 por ciento.

Conclusiones:

los métodos de cuantificación de los iones presentes en la solución concentrada para hemodiálisis con bicarbonato, permiten obtener resultados fiables en las condiciones analíticas practicadas en esta investigación(AU)
ABSTRACT

Objective:

to evaluate the analytical performance of the sodium, potassium, chloride, calcium and magnesium ions quantification methods, in a concentrated solution with bicarbonate for hemodyalisis.

Methods:

the five analytical methods are reported on the official British pharmacopeia. The performance evaluation was made by verifying linearity, accuracy and precision. In all cases, the enriched placebo method was used, with concentration levels from 60 to 140 per cent of the standard concentration value of each ion.

Results:

the five methods showed linearity in the evaluated concentration range; they were accurate and precise with global variation coefficients lower than 1.75 per cent, and a recovery rate of 100 ± 3 per cent.

Conclusions:

the quantification methods for the ions present in concentrated solution with bicarbonate for hemodialysis, allows obtaining reliable results under the analytical conditions used in this research(AU)

Texto completo: Disponível Coleções: Bases de dados nacionais / Cuba Base de dados: CUMED Tipo de estudo: Ensaio clínico controlado Idioma: Espanhol Revista: Rev. cuba. farm Ano de publicação: 2013 Tipo de documento: Artigo Instituição/País de afiliação: Laboratorio Farmacéutico Oriente/Cuba