Comparación entre dimensiones teóricas y reales del stent intracoronario en lesiones no complejas / Comparison between theoretical an actual intracoronary stent dimensions in non-complex lesions
Lozano, Iñigo; López-Palop, Ramón; Pinar, Eduardo; Pérez-Lorente, Fernando; Picó, Francisco; Valdés, Mariano.
Afiliação
  • Lozano, Iñigo; Hospital Universitario Central de Asturias. Oviedo. España
  • López-Palop, Ramón; Hospital Universitario Virgen de la Arrixaca. Murcia. España
  • Pinar, Eduardo; Hospital Universitario Virgen de la Arrixaca. Murcia. España
  • Pérez-Lorente, Fernando; Hospital Universitario Virgen de la Arrixaca. Murcia. España
  • Picó, Francisco; Hospital Universitario Virgen de la Arrixaca. Murcia. España
  • Valdés, Mariano; Hospital Universitario Virgen de la Arrixaca. Murcia. España
Rev. esp. cardiol. (Ed. impr.) ; 59(6): 624-627, jun. 2006. tab
Artigo em Es | IBECS | ID: ibc-048559
Biblioteca responsável: ES1.1
Localização: ES1.1 - BNCS
RESUMEN
El diámetro luminal mínimo intra-stent es predictor de reestenosis. Las dimensiones suministradas por el fabricante son el resultado de pruebas in vitro. El objetivo del trabajo es comparar las dimensiones reales, mediante angiografía y ultrasonidos, con las teóricas en una cohorte prospectiva de 100 lesiones no complejas susceptibles a priori de stenting directo. Se encontraron diferencias significativas entre los diámetros teóricos y reales por angiografía y ultrasonidos intracoronarios; la relación diámetro real/teórico por angiografía fue de 0,83 ± 0,09 y por ultrasonidos intracoronarios, 0,78 ± 0,10. Las medidas reales obtenidas en lesiones sin calcificación severa son significativamente inferiores que las teóricas. Las medidas nominales no deberían utilizarse como medida de referencia en el implante
ABSTRACT
The minimum in-stent lumen diameter is a predictor of restenosis. Stent dimensions provided by manufacturers are derived from in vitro tests. The aim of this study was to compare actual stent dimensions obtained by angiography and intracoronary ultrasound with dimensions that would be expected theoretically for a given inflation pressure in a cohort of 100 non-complex lesions suitable for direct stenting. Significant differences were found between the theoretical diameters and those observed by angiography and ultrasound. The actual-to-theoretical diameter ratio was 0.83 (0.09) when measured using angiography and 0.78 (0.10), using intravascular ultrasound. In lesions without severe calcification, stent dimensions were significantly smaller than indicated by the manufacturer. Nominal figures should not be used as reference values for stent implantation
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Angioplastia Coronária com Balão / Stents Limite: Humanos Idioma: Espanhol Revista: Rev. esp. cardiol. (Ed. impr.) Ano de publicação: 2006 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Central de Asturias/España / Hospital Universitario Virgen de la Arrixaca/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Angioplastia Coronária com Balão / Stents Limite: Humanos Idioma: Espanhol Revista: Rev. esp. cardiol. (Ed. impr.) Ano de publicação: 2006 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Central de Asturias/España / Hospital Universitario Virgen de la Arrixaca/España
...