Determinación indirecta y no invasiva de la presión arterial pulmonar / Noninvasive, indirect measurement of pulmonary artery pressure
Arch. bronconeumol. (Ed. impr.)
; 43(5): 267-271, mayo 2007. ilus, tab
Artigo
em Es
| IBECS
| ID: ibc-055662
Biblioteca responsável:
ES1.1
Localização: ES1.1 - BNCS
RESUMEN
Objetivo:
El procedimiento de referencia para determinar la presión arterial pulmonar (PAP) es el cateterismo derecho. Clínicamente se usa la ecocardiografía, que presenta limitaciones en pacientes con neumopatía, por mala ventana ecográfica. El objetivo del presente estudio ha sido determinar la PAP en un grupo de personas sanas mediante un procedimiento indirecto e incruento que se basa en la curva de velocidad del retorno venoso. Material ymétodos:
Se evaluó a 23 voluntarios sanos, con una edad media ± desviación estándar de 34 ± 9 años. Las medidas se realizaron con un ecógrafo Hitachi EUB 5000 (sonda de 3,5 MHz). El procedimiento consiste en situar al sujeto en decúbito supino con el cuello hiperextendido y lateralizado. Se localiza en modo B la vena yugular interna 1 cm por encima de la unión con la subclavia. Con la Doppler color se sigue dicho trayecto y con la Doppler de onda pulsada se obtienen las curvas de velocidades. Tras 5 ciclos idénticos se procede a estimar la PAP a través de la relación de velocidades máximas diastólica y sistólica.Resultados:
En el lado derecho, el valor medio de la relación de velocidades fue de 0,50 ± 0,08, que equivale a una PAP media de 15 ± 2,4 mmHg. En el lado izquierdo la relación de velocidades fue de 0,55 ± 0,09, con una PAP media de 16 ± 2,8 mmHg. La diferencia entre las PAP medias derecha e izquierda fue de 1,3 ± 3,1 mmHg, y los valores, más elevados en el lado izquierdo (p no significativo).Conclusiones:
Los valores medios de PAP obtenidos con el procedimiento descrito están en el margen de referencia para la población sana. La ecografía Doppler venosa y transcutánea es un procedimiento útil para obtener los valores de PAP en personas sanasABSTRACT
Objective:
The gold standard for measuring pulmonary artery pressure (PAP) is right heart catheterization. However, the usefulness of this technique is limited in patients with lung disease because of a poor acoustic window. The aim of this study was to assess PAP in a group of healthy persons using a noninvasive, indirect measurement procedure derived from the venous return flow velocity wave. Material andmethods:
Twenty-three healthy volunteers with a mean (SD) age of 34 (9) years were evaluated. Measurements were made with a Hitachi EUB 5000 ultrasound device (3.5 MHz wave). The subject was placed in supine decubitus with neck hyperextended and head turned to one side. B-mode images were used to locate the point in the internal jugular vein 1 cm higher than its union with the subclavian vein. A color flow image was obtained of the vessel and a pulsed wave was then used to measure velocities. After 5 identical cycles, an estimate of PAP was derived from the ratio of maximum diastolic and systolic flow rates.Results:
On the right side, the mean (SD) ratio between flow rates was 0.50 (0.08), corresponding to a mean PAP of 15 (2.4) mm Hg. On the left side, the ratio was 0.55 (0.09), corresponding to a mean PAP of 16 (2.8) mm Hg. The nonsignificant difference between mean PAP values on the right and left was 1.3 (3.1) mm Hg, and higher values were on the left.Conclusions:
Mean PAP values obtained with the described procedure are within the normal range for a healthy population. Transcutaneous venous Doppler ultrasound is useful for estimating PAP in healthy individuals
Buscar no Google
Coleções:
Bases de dados nacionais
/
Espanha
Base de dados:
IBECS
Assunto principal:
Artéria Pulmonar
/
Ecocardiografia Doppler
/
Hipertensão Pulmonar
Tipo de estudo:
Estudo diagnóstico
Limite:
Feminino
/
Humanos
/
Masculino
Idioma:
Espanhol
Revista:
Arch. bronconeumol. (Ed. impr.)
Ano de publicação:
2007
Tipo de documento:
Artigo
Instituição/País de afiliação:
Hospital General/España
/
Universidad Autónoma de Barcelona/España