Parámetros de validación de otoemisiones con el Echocheck® / Eoaes (evoked oto acoustic emissions) validation parameters with Echocheck®
Acta otorrinolaringol. esp
; 53(3): 151-155, mar. 2002. tab
Artigo
em Es
| IBECS
| ID: ibc-10391
Biblioteca responsável:
ES1.1
Localização: ES1.1 - BNCS
RESUMEN
Objetivos:
Aunque existen trabajos que demuestran la utilidad del Echocheck® en el despistaje de la hipoacusia en neonatos, ninguno lo compara con los sistemas ILO® ni tampoco han sido cotejados los resultados que el aparato muestra mediante un código de colores con los parámetros internos que genera. Se pretende demostrar que ambas técnicas son equivalentes y válidas.Métodos:
Se estudiaron mediante otoemisiones acústicas 494 oídos de recién nacidos vivos usando el sistema Echocheck. Se clasificaron en Normales, No Válidos, Falta e Inválidos según un código de colores. Posteriormente se cotejaron con los resultados internos registrados por el aparato. Por otra parte se detectaron las otoemisiones en 135 oídos con ambos sistemas y se compararon los resultados obtenidos.Resultados:
La correlación entre los resultados externos determinados mediante el código de colores y los registros internos fue absoluta según los parámetros previamente determinados. En la comparación mediante los dos sistemas, los resultados coincidieron excepto en dos pruebas en las que faltaron las otoemisiones con el Echocheck estando presentes con el ILO.Conclusiones:
Queda demostrada la validez de la prueba con este nuevo sistema, así como la equiparación de sus resultados con los obtenidos por los sistemas ILO®, recomendándose su uso para la detección precoz universal de todos los recién nacidos vivos por la sencillez de su manejo (AU)ABSTRACT
GOALS There are several papers that support the use of Echocheck system to exclude neonatal hypoacusis, but none of them compare it with the ILO systems, neither compare the internal parameters obtained versus those from a colour coded system. We are trying to demonstrate that both techniques are equivalent and valid. METHODS:
Acoustic otoemissions were studied in four hundred and ninety four (494) newborn ears using Echocheck. They were classified as Normal, Not valid, No EOAEs and Invalid using a colour coded system. After this they were compared with the internal results registered by the system. On the other hand, otoemissions were obtained using both methods (Echocheck and ILO-88) in 135 ears and the results compared.RESULTS:
The correlation between the results obtained through the colour coded system and those obtained from the internal register was absolute for those parameters previously determined. For the comparison between both systems, results did coincide in all cases except for two, in which there were only results with ILO and not with Echocheck.CONCLUSIONS:
The validity of the test using this new system as well as the comparison of the results with those obtained from ILO systems has been demonstrated, therefore its use has been recommended for the early detection of hypoacusis in newborn due to the fact that it is simple to perform (AU)
Buscar no Google
Coleções:
Bases de dados nacionais
/
Espanha
Base de dados:
IBECS
Assunto principal:
Emissões Otoacústicas Espontâneas
/
Técnicas de Diagnóstico Otológico
Tipo de estudo:
Estudo de rastreamento
Limite:
Humanos
/
Recém-Nascido
Idioma:
Espanhol
Revista:
Acta otorrinolaringol. esp
Ano de publicação:
2002
Tipo de documento:
Artigo
Instituição/País de afiliação:
Hospital Infanta Cristina/España
/
Junta de Extremadura-Servicio Extremeño de Salud/España