Comparación de la eficacia de la diálisis entre dos sistemas de control de reinfusión en hemodiafiltración en línea / Comparison of the dialysis efficacy between two reinfusion control systems in online haemodiafiltration
Aznar Barbero, Sonia; Bel Cegarra, Ramón; Badallo Mira, M. Magdalena; Beltrán Martínez, Bruno David; Pagan Escribano, David; Blázquez Fuentes, Sergio; Ponce Palacios, Ruth; Beltrán Redondo, M. Isabel.
Afiliação
  • Aznar Barbero, Sonia; Hospital Santa Lucía. Cartagena. España
  • Bel Cegarra, Ramón; Hospital Santa Lucía. Cartagena. España
  • Badallo Mira, M. Magdalena; Hospital Santa Lucía. Cartagena. España
  • Beltrán Martínez, Bruno David; Hospital Santa Lucía. Cartagena. España
  • Pagan Escribano, David; Hospital Santa Lucía. Cartagena. España
  • Blázquez Fuentes, Sergio; Hospital Santa Lucía. Cartagena. España
  • Ponce Palacios, Ruth; Hospital Santa Lucía. Cartagena. España
  • Beltrán Redondo, M. Isabel; Hospital Santa Lucía. Cartagena. España
Enferm. nefrol ; 15(3): 182-187, jul.-sept. 2012. tab, ilus
Artigo em Espanhol | IBECS | ID: ibc-107461
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN
La dosis de diálisis y volumen de reinfusión son elementos claves en los mejores resultados comunicados en la hemodiafiltración on-line. Avances tecnológicos permiten dos formas de control automatizado del volumen de reinfusión (una volumétrica y otra por presión), y dos métodos de medición del Kt obtenido por dialisancia iónica (OCM y Diascan). El propósito del presente estudio fue compararlos. Se diseña un estudio prospectivo cruzado sobre población prevalente en hemodiafiltración on-line posdilucional, con consentimiento expreso. Durante 2 semanas consecutivas se realizan tres sesiones con reinfusión automática según hematocrito y proteínas totales, y medición del Kt por OCM, y tres sesiones con reinfusión automática en función de la presión transmembrana y medición del Kt por Diascan. Se compara el valor promedio de volumen de reinfusión y Kt, contrastando las variables con la t-student para muestras relacionadas, estableciendo un valor de significación estadística p < 0,05. Dieciséis pacientes concluyen el estudio, 56% hombres, 58,8 años. Un 81,3% presentan fístula autóloga o protésica como acceso vascular. El volumen de reinfusión es significativamente mayor (p = 0,029) bajo control de presión (23,24 ± 2,55 l versus 21,81 ± 1,75 l) mientras que el Kt medido por OCM es significativamente (p < 0,001) mayor que con Diascan (59,94 ± 5,05 l versus 55,12 ± 4,15 l). Encontramos un incremento medio del volumen de reinfusión (6,2%) bajo control de presión, si bien ambos métodos obtienen buenos resultados en los litros de trasporte convectivo final. La dosis de diálisis administrada (Kt) es mayor utilizando OCM (AU)
ABSTRACT
The dose of dialysis and the volume of reinfusión are key elements in the best results communicated in the hemodiafiltration online. Technological advances allow two forms of automated control of reinfusion volume (a volume-based and a pressure-based control system), and two methods of measuring Kt obtained by ionic dialysance (OCM and Diascan). The purpose of this study was to compare both these control systems. We designed a prospective, crossover study of the prevalent population in post-dilution on-line hemodiafiltration, with explicit consent. For 2 consecutive weeks, three sessions were performed with automatic reinfusion according to hematocrit and total protein values and Kt measurement with OCM, and three sessions were performed with automatic reinfusion according to transmembrane pressure and Kt measurement with Diascan. The average reinfusion volume and Kt were compared. Variables were contrasted with Student’s t-test for related samples. Statistical significance was set at p <0.05. Sixteen patients concluded the study (56% men, with a mean age of 58.8 years). A total of 81.3% had autologous or prosthetic fistulas for vascular access. Reinfusion volume was significantly higher (p = 0.029) under pressure control (23.24 ± 2.55 lversus 21.81 ± 1.75 l) while Kt measured by OCM was significantly (p <0.001) higher than that measured by Diascan (59.94 ± 5.05 l versus 55.12 ± 4.15 l. We found a 6.2% increase in volume reinfusion under pressure control, although both methods performed well in the final convective transport liters. The dialysis dose (Kt) was higher when OCM was used (AU)
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Diálise Renal / Resultado do Tratamento / Hemodiafiltração / Avaliação de Eficácia-Efetividade de Intervenções / Recursos Humanos de Enfermagem Tipo de estudo: Estudo observacional Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Enferm. nefrol Ano de publicação: 2012 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Santa Lucía/España

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Diálise Renal / Resultado do Tratamento / Hemodiafiltração / Avaliação de Eficácia-Efetividade de Intervenções / Recursos Humanos de Enfermagem Tipo de estudo: Estudo observacional Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Enferm. nefrol Ano de publicação: 2012 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Santa Lucía/España
...