Your browser doesn't support javascript.
loading
Evaluación de un nuevo reactivo monoclonal, N-Látex Free Light Chains, para la cuantificación nefelométrica de las cadenas ligeras Kappa y Lambda libres en suero / Evaluation of a new monoclonal reagent, N-Látex Free Light Chains, for nefelometric quantification of free light chain kappa and lambda in serum
Máiz Suárez, Lola; Garnacho Gayarre, Nuria; Cabo del Riego, Julia; Penedo Pita, Manuel; Formoso Lavandeira, Dolores; Rueda Rúa, Ramiro.
Afiliação
  • Máiz Suárez, Lola; Hospital Universitario Lucus Augusti (HULA). Lugo. España
  • Garnacho Gayarre, Nuria; Hospital Universitario Lucus Augusti (HULA). Lugo. España
  • Cabo del Riego, Julia; Hospital Universitario Lucus Augusti (HULA). Lugo. España
  • Penedo Pita, Manuel; Hospital Universitario Lucus Augusti (HULA). Lugo. España
  • Formoso Lavandeira, Dolores; Hospital Universitario Lucus Augusti (HULA). Lugo. España
  • Rueda Rúa, Ramiro; Hospital Universitario Lucus Augusti (HULA). Lugo. España
Rev. lab. clín ; 6(1): 18-25, ene.-mar. 2013.
Article em Es | IBECS | ID: ibc-110352
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN
Introducción. Con la evaluación del ensayo de cadenas ligeras libres en suero Kappa y Lambda (FLC), pretendemos colaborar en la imprescindible estandarización internacional de esta técnica. Su utilidad en el diagnóstico y seguimiento de mielomas, gammapatías de significado incierto (MGUS), amiloidosis, plasmocitomas solitarios etc. está ampliamente demostrada en la bibliografía, las guías clínicas internacionales las incluyen para el específico diagnóstico de las discrasias de células plasmáticas. Material y métodos. Se evaluó un nuevo reactivo N-Látex CLL para nefelometría, basado en anticuerpos monoclonales, que lleva adjunto un reactivo suplementario específico, que fue desarrollado con inmunoglobulinas de ratón. Resultados. La exactitud y precisión del método evaluadas con controles de 3 niveles de concentración fueron aceptables. Se ha comparado este reactivo con el Freelite(R), procesando 30 muestras de pacientes, por el nefelómetro BNII (Siemens), se solventaron los falsos positivos del 2.°. Así mismo, se comparó el método por nefelometría (BN II) frente a turbidimetría (SPA PLUS) los resultados fueron favorables a la nefelometría. En cuanto a sensibilidad y especificidad del método evaluado, se realizaron las curvas ROC, con 60 pacientes y se obtuvieron resultados con una moderada correlación. Discusión. Consideramos que la falta de reproductibilidad en la cuantificación de las FLC existente, se debía a los factores: alta variabilidad entre los lotes de reactivos de la misma casa comercial; diferencias en los resultados de las muestras con elevadas concentraciones de FLC por no incluir el chequeo de exceso de antígeno, y diferencias en los principios de medida de los autoanalizadores utilizados (AU)
ABSTRACT
Introduction. Evaluation of testing of serum free light chains Kappa and Lambda (FLC), to collaborate on the essential international standardization of this technique. Its usefulness in diagnosis and follow-up of myeloma, Gammapatías of undeterminated significance (MGUS), amyloidosis, solitary plasmocitomas etc. is amply demonstrated in the literature, the international clinical guidelines include them for the specific diagnosis of the discrasias of plasma cells. Material and methods. We evaluated a new reagent N-Látex CLL for Nephelometry, based on monoclonal antibodies, which leads attached a supplementary reagent specific, which was developed with mouse immunoglobulins. Results. The accuracy and precision of the method evaluated with 3 levels of concentration controls were acceptable. It has been compared this reagent with the existing 30 samples from patients with the nephelometer BNII (Siemens), false positives of the 2nd were overcome. Likewise compared the method by nephelometry (BN II) against turbidimetry (SPA PLUS) the results were favourable to the nephelometry reference technique. In terms of sensitivity and specificity of the evaluated method, ROC curves, with 60 patients were conducted and results with a moderate correlation. Discussion. We believe that the lack of reproducibility in the quantification of the FLC due to factors: high variability between batches of reagents in the same commercial House; differences in the results of the samples with high concentrations of FLC, for not including the checking of antigen excess, and differences in the principles of the used clinical measurement (AU)
Assuntos
Buscar no Google
Coleções: 06-national / ES Base de dados: IBECS Assunto principal: Cadeias lambda de Imunoglobulina / Mieloma Múltiplo / Nefelometria e Turbidimetria Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies / Qualitative_research Limite: Animals Idioma: Es Revista: Rev. lab. clín Ano de publicação: 2013 Tipo de documento: Article
Buscar no Google
Coleções: 06-national / ES Base de dados: IBECS Assunto principal: Cadeias lambda de Imunoglobulina / Mieloma Múltiplo / Nefelometria e Turbidimetria Tipo de estudo: Diagnostic_studies / Guideline / Prognostic_studies / Qualitative_research Limite: Animals Idioma: Es Revista: Rev. lab. clín Ano de publicação: 2013 Tipo de documento: Article