Valores de referencia en plasma de osmolalidad, electrolitos, calcio, magnesio y urea en una población del centro de España / Reference values in plasma for osmolality, electrolytes, calcium, magnesium and urea, in a spanish population
Martín Calderón, José Luis; Varona Pérez, Julia; Bustos Guadaño, Fernando; Fuente Mateo, Pilar de La; Solano Ramos, Fernando.
Afiliação
  • Martín Calderón, José Luis; Hospital General Nuestra Señora del Prado. Talavera. España
  • Varona Pérez, Julia; Hospital General Nuestra Señora del Prado. Talavera. España
  • Bustos Guadaño, Fernando; Hospital General Nuestra Señora del Prado. Talavera. España
  • Fuente Mateo, Pilar de La; Hospital General Nuestra Señora del Prado. Talavera. España
  • Solano Ramos, Fernando; Hospital General Nuestra Señora del Prado. Talavera. España
Rev. lab. clín ; 7(2): 49-54, abr.-jun. 2014. tab
Artigo em Espanhol | IBECS | ID: ibc-124999
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Introducción:

El establecimiento de los valores de referencia es una tarea esencial de los laboratorios clínicos, ya que pueden variar según los instrumentos empleados y la población considerada. En el presente trabajo pretendemos definir el intervalo de referencia en nuestra población para las siguientes magnitudes biológicas en plasma osmolalidad, sodio, potasio, calcio, urea, y magnesio con el analizador Unicell DXC 800. Material y

métodos:

Se seleccionó una muestra de referencia formada por 146 individuos sanos entre los donantes que acudieron a nuestro banco de sangre y se les extrajo sangre en un tubo de heparina litio que fue centrifugado a 1.700 g durante 5 minutos para la obtención de plasma. En el plasma resultante se analizaron urea, sodio, potasio, calcio y magnesio con el analizador Unicell DXC 800 (Beckman Coulter®), midiéndose la osmolalidad con el osmómetro 3300 (Advanced Instruments®). Se aplicó el método no paramétrico del CLSI C28-A3 para la obtención de los valores de referencia.

Resultados:

Se obtuvieron los siguientes intervalos de referencia urea (13,67-48 mg/dL), sodio 136,7-145 mmol/L), potasio (3,37-4,6 mmol/L), osmolalidad (290-305,42 mosm/kg), calcio (8-9,3 mg/dL) y magnesio (1,9-2,2 mg/dL). Estos intervalos fueron diferentes a los propuestos por el fabricante, lo que refuerza la necesidad de producir valores de referencia propios (AU)
ABSTRACT

Introduction:

The establishing of reference values is an essential task of clinical laboratories. In this work the reference intervals for the following parameters in plasma were determined osmolality, sodium, potassium, calcium, magnesium and urea, using a Unicell DXC 800 analyzer (Beckman Coulter®). Material and

methods:

A sample of 146 healthy individuals was selected from health blood donors of our institution. Plasma was obtained by centrifugation of lithium heparin tubes at 1700 g for five minutes. Sodium, potassium, calcium, magnesium and urea were measured on a Unicell DXC 800 analyzer (Beckman Coulter®), whereas osmolality was measured with an osmometer (Advanced Instruments® 3300). The CLSI C28 A-3 non-parametric method was used to determine the reference intervals.

Results:

The reference intervals obtained were osmolality (290-305.42 mosm/kg), sodium (136.7-145 mmol/L), potassium (3.37-4.6 mmol/L), calcium (8-9.3 mg/dL), magnesium (1.9- 2.2 mg/dL) and urea (13.67-48 mg/dL). These values are different from those claimed by the manufacturer, which reinforces the need for determining individual reference values in each laboratory (AU)
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Concentração Osmolar / Ureia / Cálcio / Eletrólitos / Magnésio Limite: Humanos Idioma: Espanhol Revista: Rev. lab. clín Ano de publicação: 2014 Tipo de documento: Artigo Instituição/País de afiliação: Hospital General Nuestra Señora del Prado/España

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Concentração Osmolar / Ureia / Cálcio / Eletrólitos / Magnésio Limite: Humanos Idioma: Espanhol Revista: Rev. lab. clín Ano de publicação: 2014 Tipo de documento: Artigo Instituição/País de afiliação: Hospital General Nuestra Señora del Prado/España
...