Concordancia entre la tonometría de Goldmann, Icare Pro y Corvis ST / Comparison between Goldmann, Icare Pro and Corvis ST tonometry
Arch. Soc. Esp. Oftalmol
; 89(7): 260-264, jul. 2014. ilus, graf
Artigo
em Espanhol
| IBECS
| ID: ibc-125818
Biblioteca responsável:
ES1.1
Localização: BNCS
RESUMEN
OBJETIVO:
Comparar los valores de presión intraocular (PIO) del nuevo tonómetro de no contacto Corvis ST (CST) con la tonometría de aplanación de Goldmann (GAT) y la tonometría de rebote Icare Pro (PRO). MATERIAL YMÉTODOS:
Se seleccionaron 178 ojos de 178 voluntarios sanos a los que se les midió la PIO de manera aleatorizada con CST, PRO y GAT. Se midió el espesor corneal central (GCC) con paquimetría ultrasónica. Para cada variable, se tomaron 3 medidas y la media de las 3 fue introducida en el análisis estadístico. La concordancia entre los tonómetros se determinó mediante las gráficas de Bland-Altmann.RESULTADOS:
La PIO medias obtenidas con los 3 tonómetros fueron de 15,5 ± 2,8 mmHg para GAT, 14,6 ± 2,3 mmHg para PRO y 15,4 ± 2,8 mmHg para CST. La diferencia media entre pares de tonómetros fue GAT-PRO = 0,9 ± 1,7 mmHg (p < 0,001); GAT-CST 0,1 ± 2,2 (p = 0,398) y PRO-CST −0,8 ± −0,7 mmHg, p < 0,001. En los 3 casos se encontró una correlación significativa con el GCC GAT r = 0,325, p < 0,001; PRO r = 0,385, p < 0,001 y CST r = 0,428; p < 0,001.CONCLUSIONES:
Las diferencias encontradas entre el PRO y el GAT son significativamente mayores que entre el CST y el GAT, sin diferencias significativas entre las medidas de estos 2. Las medidas de los 3 tonómetros se ven afectadas por el GCCABSTRACT
PURPOSE:
To compare intraocular pressure (IOP) between the new non-contact tonometer Corvis ST (CST), the Goldmann applanation tonometry (GAT) and Icare Pro rebound tonometer (PRO).METHODS:
A total of 178 eyes of 178 healthy subjects were selected for the study. Measurements of IOP were made in a random order with GAT, PRO and CST. Central corneal thickness (CCT) was determined by ultrasound pachymetry. The mean of three valid measurements of each variable was used in the statistical analysis. The relationship between the tonometers was established using Bland-Altman plots.RESULTS:
Mean IOP was 15.5±2.8 mmHg for GAT, 15.4±2.8 mmHg for CST, and 14.6±2.3 mmHg for PRO. The mean differences between pairs of tonometers were GATPRO= 0.9±1.7 mmHg (P<0.001), GAT-CST 0.1±2.2 (P=0.398), and PRO-CST −0.8±−0.7 mmHg,p < 0.001. A positive relationship was detected between CCT and the three tonometers GATr = 0.325, P<0.001; PRO r = 0.385, P<0.001, and CST r = 0.428; P<0.001.CONCLUSIONS:
The differences found between PRO and GAT were significantly higher than those found between CST and GAT, which showed non-significant differences. The measurements of the three tonometers were affected by the CCT
Buscar no Google
Coleções:
Bases de dados nacionais
/
Espanha
Base de dados:
IBECS
Assunto principal:
Tonometria Ocular
/
Determinação da Pressão Arterial
/
Pressão Intraocular
Tipo de estudo:
Ensaio clínico controlado
Limite:
Humanos
Idioma:
Espanhol
Revista:
Arch. Soc. Esp. Oftalmol
Ano de publicação:
2014
Tipo de documento:
Artigo
Instituição/País de afiliação:
Instituto de Investigación Sanitaria del Hospital Clínico San Carlos/España