Artifacts in brain magnetic resonance imaging due to metallic dental objects
Med. oral patol. oral cir. bucal (Internet)
; 14(6): e278-e282, jun. 2009. tab, ilus
Artigo
em Inglês
| IBECS
| ID: ibc-136367
Biblioteca responsável:
ES1.1
Localização: BNCS
ABSTRACT
Objective:
Artifacts caused by metallic objects, such as dental crowns, dental implants and metallic orthodontic appliances, are a common problem in head and neck MRI. The aim of this retrospective study was to identify the main metal dental objects that produce artifacts on brain MRIs. Studydesign:
Imaged metallic artifacts and their sources were identified. Artifact image plane was rated on a score of 0 or 1 (0 - distinguishable for diagnosis and 1 - not distinguishable for diagnosis).Results:
Seventy-eight percent of the artifacts appearing in images were caused by orthodontic appliances, followed by dental titanium implants (18%) and metallic crowns (4%). Orthodontic appliances obtained the highest scores in all planes.Conclusions:
We concluded that is difficult to avoid the effect of metallic artifacts in the maxillofacial regions on brain scan. Removing metallic parts of the orthodontic appliance should ensure diagnostically useful quality scans (AU)
Buscar no Google
Coleções:
Bases de dados nacionais
/
Espanha
Base de dados:
IBECS
Assunto principal:
Imageamento por Ressonância Magnética
/
Artefatos
/
Materiais Dentários
Tipo de estudo:
Estudo observacional
/
Estudo prognóstico
Limite:
Humanos
Idioma:
Inglês
Revista:
Med. oral patol. oral cir. bucal (Internet)
Ano de publicação:
2009
Tipo de documento:
Artigo
Instituição/País de afiliação:
University of Campinas/Brazil