Comparación entre dos inmunoensayos automatizados por quimioluminiscencia para la cuantificación de 25(OH) vitamina D / Comparison between two automated chemiluminescence immunoassays for quantifying 25 (OH) vitamin D
Torrubia, B; Alonso, I; López-Ramiro, E; Mahillo, I; Piedra, C de la.
Afiliação
  • Torrubia, B; Fundación Jiménez Díaz. Laboratorio de Bioquímica. Madrid. España
  • Alonso, I; Fundación Jiménez Díaz. Laboratorio de Bioquímica. Madrid. España
  • López-Ramiro, E; Fundación Jiménez Díaz. Laboratorio de Bioquímica. Madrid. España
  • Mahillo, I; Fundación Jiménez Díaz. Servicio de Epidemiología. Madrid. España
  • Piedra, C de la; Fundación Jiménez Díaz. Laboratorio de Bioquímica. Madrid. España
Rev. osteoporos. metab. miner. (Internet) ; 8(2): 70-74, abr.-jun. 2016. graf
Artigo em Espanhol | IBECS | ID: ibc-154853
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Introducción:

La cuantificación de 25(OH) vitamina D total en sangre es el marcador más preciso del estado de vitamina D en un individuo. La técnica patrón-oro para su medición es la cromatografía líquida/ espectrometría de masas en tándem (LC-MS/MS), aunque actualmente los laboratorios clínicos utilizan de rutina técnicas de quimioluminiscencia. El objetivo del estudio fue comparar las concentraciones de 25(OH) vitamina D obtenidas mediante dos métodos automatizados comerciales y estudiar la correlación de dichos métodos con la técnica de referencia LC-MS/MS. Material y

método:

Se cuantificaron los niveles de 25(OH) vitamina D en 1.000 muestras de suero del laboratorio de Bioquímica de la Fundación Jiménez Díaz mediante 2 métodos automatizados comerciales por detección de quimioluminiscencia ADVIA CENTAURO® (SIEMENS) y LUMIPULSE® G1200 (FUJIREBIO). Entre todas las muestras analizadas, las 50 más discordantes entre sí se enviaron para ser evaluadas por la técnica de referencia LC-MS/MS.

Resultados:

Los resultados obtenidos indican que existe una buena correlación entre los dos

métodos:

CCI=0,923 (0,914-0,932), siendo los valores de LUMIPULSE® G1200 un 10% superiores a los de CENTAURO ®. Con respecto a las 50 muestras seleccionadas, podemos observar que existe una buena correlación entre los dos inmunoensayos con la técnica LC-MS/MS, aunque ambos métodos infraestiman considerablemente los resultados de 25(OH) vitamina D con respecto al patrón-oro.

Discusión:

Aunque ambas técnicas son adecuadas para su utilización, habría que plantearse si la «epidemia» mundial de hipovitaminosis D se debe a la metodología de análisis utilizada. Esta variabilidad entre inmunoensayos se solucionaría estandarizando las diferentes técnicas comerciales con los materiales de referencia elaborados por el NIST (AU)
ABSTRACT

Introduction:

Quantifying total blood 25 (OH) vitamin D is the most accurate marker of an individual’s vitamin D status. The gold standard technique for measurement is liquid chromatography tandem mass spectrometry (LC-MS/MS), although currently clinical laboratories tend to use chemiluminescence techniques. The objective of this study was to compare 25 (OH) vitamin D concentrations obtained by two commercially-produced automated methods and study the correlation of these methods with the LC-MS/MS reference technique. Material and

methods:

The 25(OH) vitamin D levels were quantified in 1,000 serum samples from the Jimenez Diaz Biochemistry Foundation Laboratory using 2 automated methods for chemiluminescence detection ADVIA CENTAURO® (SIEMENS) and LUMIPULSE® G1200 (Fujirebio). Among all the samples tested, the 50 most discordant to each other were sent to be evaluated by LC-MS/MS reference technique.

Results:

The results indicate that there is good correlation between the two

methods:

CCI=0.923 (0.914-0.932), with the G1200 LUMIPULSE® values 10% being higher than CENTAURO®. Regarding the 50 samples selected, we can see that there is a good correlation between the two immunoassays with LC-MS/MS, although both methods significantly underestimate 25 (OH) vitamin D results with respect to the gold standard.

Discussion:

Although both techniques are suitable for use, it is worth considering whether the worldwide vitamin D deficiency epidemic is due to the analysis methodology used. This variability between immunoassays could be solved by standardizing the different commercial techniques in line with NIST produced reference materials (AU)
Assuntos
Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Vitamina D / Cromatografia Líquida / Medições Luminescentes Tipo de estudo: Estudo diagnóstico Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. osteoporos. metab. miner. (Internet) Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: Fundación Jiménez Díaz/España
Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Vitamina D / Cromatografia Líquida / Medições Luminescentes Tipo de estudo: Estudo diagnóstico Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. osteoporos. metab. miner. (Internet) Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: Fundación Jiménez Díaz/España
...