Rotura por fatiga de la zona central del vástago Exeter(R): una complicación excepcional / Fatigue fracture of the central zone of Exeter stem: an unusual complication
Barrionuevo Sánchez, FJ; Rial Valverde, L; Villa Gil-Ortega, M.
Afiliação
  • Barrionuevo Sánchez, FJ; Hospital San Juan de Dios del Aljarafe. Servicio de Cirugía ortopédica y traumatología. Bormujos. España
  • Rial Valverde, L; Hospital San Juan de Dios del Aljarafe. Servicio de Cirugía ortopédica y traumatología. Bormujos. España
  • Villa Gil-Ortega, M; Hospital San Juan de Dios del Aljarafe. Servicio de Cirugía ortopédica y traumatología. Bormujos. España
Rev. Soc. Andal. Traumatol. Ortop. (Ed. impr.) ; 33(3): 57-62, jul.-sept. 2016. ilus
Artigo em Espanhol | IBECS | ID: ibc-157839
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Objetivos:

Conocer una complicación rara pero con importante morbimortalidad como es la rotura de vástagos protésicos cementados. Material y

métodos:

En nuestro caso nos referimos a la rotura de un vástago pulido Exeter(R) (Stryker(R)), fabricado en acero inoxidable y ampliamente utilizado en todo el mundo. Tras implantar dicho vástago en una paciente de 74 años que sufrió una fractura subcapital, se observa a los 8 años de la cirugía una rotura del vástago en la zona central; por lo que se recambia dicho vástago por uno de anclaje diafisario con buena evolución. Resultados y

conclusiones:

tras analizar las posibles causas que generaron el fracaso del vástago, identificamos un defecto en la cementación como el principal causante del fallo del implante
ABSTRACT
Objetives explain a rare but important complication for patient that is the breakage of cemented stem.

Methods:

in our case, we refer to a polished stem Exeter(R) (Stryker(R)), made of stainless steel and widely used around the world. We implant Exeter® stem in a 74 year old patient and 8 years after the surgery, is observed a rupture in the central area of the stem; that stem is exchanged for one of diaphyseal anchorage with good evolution.

Results:

after analyzing the possible causes of the failure of this stem, we identified a defect in the cement as the main cause of implant failure
Assuntos
Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Próteses e Implantes / Terapêutica / Radiografia / Fraturas de Estresse / Fraturas do Fêmur / Prótese de Quadril Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Limite: Idoso / Feminino / Humanos Idioma: Espanhol Revista: Rev. Soc. Andal. Traumatol. Ortop. (Ed. impr.) Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: Hospital San Juan de Dios del Aljarafe/España
Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Próteses e Implantes / Terapêutica / Radiografia / Fraturas de Estresse / Fraturas do Fêmur / Prótese de Quadril Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Limite: Idoso / Feminino / Humanos Idioma: Espanhol Revista: Rev. Soc. Andal. Traumatol. Ortop. (Ed. impr.) Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: Hospital San Juan de Dios del Aljarafe/España
...