Valoración del ángulo de la cámara anterior mediante biomicroscopia ultrasónica, gonioscopia y examen de Van Herick / Evaluation of the angle of the anterior chamber using ultrasound biomicroscopy, gonioscopy and a Van Herick examination
Carrizosa Murcia, M; Rey Rodríguez, DV.
Afiliação
  • Carrizosa Murcia, M; Universidad El Bosque. Ciencias de la Visión. Bogotá. Colomiba
  • Rey Rodríguez, DV; Universidad El Bosque. Epidemología. Bogotá. Colomiba
Arch. Soc. Esp. Oftalmol ; 94(3): 114-118, mar. 2019. tab, graf
Artigo em Espanhol | IBECS | ID: ibc-178311
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Objetivo:

Comparar la medición del ángulo de la cámara anterior mediante biomicroscopia ultrasónica (UBM), gonioscopia y examen de Van Herick.

Métodos:

Se realizó un estudio comparativo observacional entre 3 métodos diferentes para la estimación del ángulo iridocorneal. Fueron evaluados 30 sujetos con ángulos abiertos y cerrados a través del examen con UBM, gonioscopia y Van Herick. En la valoración con UBM y gonioscopia se incluyeron en el análisis todos los cuadrantes; en la estimación de Van Herick se analizaron los cuadrantes nasal y temporal. Todos los exámenes incluyeron la valoración de ambos ojos en condiciones de iluminación fotópica y escotópica. Estos hallazgos fueron comparados con la prueba de concordancia kappa.

Resultados:

No hubo diferencia estadísticamente significativa entre los 3métodos para evaluar el estado del ángulo iridocorneal. La consistencia entre UBM y gonioscopia fue mejor (kappa 0,93) con relación a la estimación del ángulo con Van Herick (kappa 0,83).

Conclusiones:

Los exámenes UBM, gonioscopia y Van Herick tienen resultados comparables. La gonioscopia en primera instancia estaría indicada como prueba de rutina para identificación de cierre angular
ABSTRACT

Objective:

To compare the measurement of the angle of the anterior chamber using ultrasound biomicroscopy (UBM), gonioscopy and the Van Herick examination.

Methods:

An observational comparative study was carried out between 3different methods for the estimation of the iridocorneal angle. A total of 30 subjects with open and closed angles were evaluated using UBM, gonioscopy and Van Herick. In the UBM and gonioscopy assessment all the quadrants were included in the analysis, and in the Van Herick examination the nasal and temporal quadrant were also analysed. All the examinations included the assessment of both eyes under photo-optical and scotopic illumination conditions. These findings were compared with the kappa concordance test.

Results:

There was no statistically significant difference between the 3 methods in the evaluation of the state of the iridocorneal angle. The consistency between UBM and gonioscopy was better (kappa 0.93) in relation to the estimation of the angle with Van Herick (kappa 0.83).

Conclusions:

The UBM, gonioscopy and Van Herick examinations have comparable results, although gonioscopy in the first instance would be indicated as a routine test for identification of angular closure
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Microanálise por Sonda Eletrônica / Câmara Anterior Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Arch. Soc. Esp. Oftalmol Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Universidad El Bosque/Colomiba
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Microanálise por Sonda Eletrônica / Câmara Anterior Limite: Adulto / Feminino / Humanos / Masculino Idioma: Espanhol Revista: Arch. Soc. Esp. Oftalmol Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Universidad El Bosque/Colomiba
...