Eficacia y seguridad de la mascarilla laríngea Protector (R) en procedimientos respiratorios invasivos / Efficacy and safety of the Protector (R) laryngeal mask in invasive respiratory procedures
Ortega Mera, U; González Larrocha, O; Flores de la Rica, M.
Afiliação
  • Ortega Mera, U; Hospital de Galdakao-Usansolo. FEA Anestesia y Reanimación. Bizkaia. España
  • González Larrocha, O; Hospital de Galdakao-Usansolo. FEA Anestesia y Reanimación. Bizkaia. España
  • Flores de la Rica, M; Hospital de Galdakao-Usansolo. Servicio de Anestesia. Bizkaia. España
Cir. mayor ambul ; 26(4): 190-194, 2021. tab, ilus
Artigo em Espanhol | IBECS | ID: ibc-206751
Biblioteca responsável: ES1.1
Localização: ES15.1 - BNCS
RESUMEN

Objetivos:

Las pruebas respiratorias invasivas pueden ser realizadas bajo sedación profunda o preferentemente bajo anestesia general para facilitar su rea- lización y evitar complicaciones. Dichas pruebas requieren del manejo compartido de la vía aérea junto al neumólogo, por lo que tubos endotraqueales de gran calibre será necesarios para garantizar la ventilación. Una alternativa a la intubación sería la colocación de una mascarilla laríngea (ML). Pocos estudios recogen la realización de dichas pruebas con ML y no existe ningún estudio en el que se haya realizado con la ML Protector®. Esta ML presenta un gran canal calibre interno y la suficiente flexibilidad para poder realizarla reduciendo los riesgos de intubación. El objetivo de este estudio es demostrar la seguridad de la mascarilla laríngea Protector® para la realización de pruebas respiratorias invasivas.

Metodología:

Se reclutaron 55 pacientes consecutivos que fueron sometidos a pruebas respiratorias invasivas con la mascarilla laríngea Protector®.

Resultados:

La tasa de éxito con la ML Protector® ha sido del 92 % sin presentar complicaciones. La ventilación es garantizada con menores presiones pico que con un tubo endotraqueal, ningún paciente ha referido dolor de garganta y todos los pacientes fueron dados de alta como máximo a las 4 horas tras el procedimiento.

Conclusiones:

La realización de pruebas respiratorias invasivas con ML Protector® es seguro reduciendo las complicaciones asociadas a intubaciones con tubos de gran calibre y reduciendo las presiones para asegurar su ventilación. (AU)
ABSTRACT

Objective:

Invasive respiratory tests can be performed under deep sedation or preferably under general anesthesia to facilitate their performance and avoid complications. These procedures require shared management of the airway with the pulmonologist, so large-caliber endotracheal tubes will be necessary to guarantee ventilation. An alternative to intubation would be the placement of a laryngeal mask. Few studies record the performance of such tests with LM and there are no studies in which it has been performed with LM Protector®. This LM presents a large internal caliber canal and sufficient flexibility to perform it, reducing the risks of intubation. The objective of this study is to demonstrate the safety of the Protector® laryngeal mask for invasive respiratory procedures.

Method:

Fifty-five consecutive patients who underwent invasive respiratory procedures with the Protector® laryngeal mask were recruited.

Results:

The success rate with the LM Protector has been 92 % without complications. Ventilation is guaranteed with lower peak pressures, no patient has reported a sore throat and all patients were discharged at most 4 hours after the procedure.

Conclusions:

Performing ventilatory respiratory procedures with LM Protector® is safe, reducing the complications associated with large-caliber tubes intubation. (AU)
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Pneumologia / Eficácia / Máscaras Laríngeas Limite: Humanos Idioma: Espanhol Revista: Cir. mayor ambul Ano de publicação: 2021 Tipo de documento: Artigo Instituição/País de afiliação: Hospital de Galdakao-Usansolo/España
Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Pneumologia / Eficácia / Máscaras Laríngeas Limite: Humanos Idioma: Espanhol Revista: Cir. mayor ambul Ano de publicação: 2021 Tipo de documento: Artigo Instituição/País de afiliação: Hospital de Galdakao-Usansolo/España
...