Concordance between the Piko - 1(R) portable device and pneumotachography in measuring PEF and FEV1 in asthmatic children
Allergol. immunopatol
; 37(5): 244-248, sept.-oct. 2009. ilus, graf
Artigo
em Inglês
| IBECS
| ID: ibc-73387
Biblioteca responsável:
ES1.1
Localização: BNCS
ABSTRACT
Objective:
To assess concordance in the measurement of peak expiratory flow (PEF) and forced expiratory volume in one second (FEV1) between the portable device Piko-1 (Ferraris) and a pneumotachograph. Patients andmethods:
Forced spirometry (Master Screen Jaeger) was performed according to ATS/ERS norms, selecting the best value of three curves, and three measurements with the Piko-1 were recorded, following the recommendations of the manufacturer. Results. Eighty patients between 518 years of age were studied. Based on the Bland-Altman method, the mean differences obtained were 9.82 (95%CI 2.4317.21) for PEF and 0.17 (95%CI 0.120.21) for FEV1. The intraclass correlation coefficient was 0.96 (p<0,001; 95%CI 0.930.97) for FEV1 and 0.93 (p<0,001; 95%CI 0.890.95) for PEF.Conclusions:
Piko-1 offers FEV1 measurements close to those obtained with forced spirometry, thus allowing more exact patient assessment in home-based follow-up, emergency services, or hospital wards (AU)RESUMEN
No disponible
Buscar no Google
Coleções:
Bases de dados nacionais
/
Espanha
Base de dados:
IBECS
Assunto principal:
Asma
/
Espirometria
/
Volume Expiratório Forçado
/
Fluxo Expiratório Máximo
Limite:
Adolescente
/
Criança
/
Criança, pré-escolar
/
Feminino
/
Humanos
/
Masculino
Idioma:
Inglês
Revista:
Allergol. immunopatol
Ano de publicação:
2009
Tipo de documento:
Artigo
Instituição/País de afiliação:
Hospital Infantil Universitario Niño Jesús/Spain
/
Hospital Universitario Materno Infantil de Canarias/Spain