Validación de la medición transcutánea de la presión parcial de dióxido de carbono en el paciente crítico pediátrico / Validation of the transcutaneous carbon dioxide tension measurements in critical paediatric patients
Fernández de Miguel, S; Gaboli, M; González-Celador, R; Gómez de Quero, P; Murga Herrero, V; Sánchez Granados, J. M; Payo Pérez, R.
Afiliação
  • Fernández de Miguel, S; Hospital Universitario de Salamanca. Salamanca. España
  • Gaboli, M; Hospital Universitario de Salamanca. Salamanca. España
  • González-Celador, R; Hospital Universitario de Salamanca. Salamanca. España
  • Gómez de Quero, P; Hospital Universitario de Salamanca. Salamanca. España
  • Murga Herrero, V; Hospital Universitario de Salamanca. Salamanca. España
  • Sánchez Granados, J. M; Hospital Universitario de Salamanca. Salamanca. España
  • Payo Pérez, R; Hospital Universitario de Salamanca. Salamanca. España
An. pediatr. (2003, Ed. impr.) ; 72(3): 165-171, mar. 2010.
Artigo em Espanhol | IBECS | ID: ibc-78509
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Objetivo:

Valorar la fiabilidad de la medición de la presión transcutánea de dióxido de carbono (PtCO2) respecto a la medición de la presión arterial de dióxido de carbono (PaCO2). Material y

métodos:

Estudio analítico, observacional, longitudinal y prospectivo. Cohorte de pacientes ingresados en unidad de cuidados intensivos pediátricos. La PtCO2 se midió con el monitor digital SenTec, aplicando el sensor con un anillo específico (sensor V-sign, versión MDB 04.04.02). Se recogieron al mismo tiempo la PtCO2 y PaCO2. La significación estadística de la asociación se calculó mediante el test F de Snedecor, el coeficiente de correlación r2 de Pearson y el coeficiente de correlación intraclase. El grado de acuerdo se estimó con el método de Bland y Altman. La consistencia de los resultados se estudió con el ANOVA.

Resultados:

Se compararon 106 mediciones pareadas de PtCO2 y PaCO2, de 12 pacientes. Las PaCO2 y PtCO2 medias fueron 51,0±13mmHg y 50,1±14mmHg; r2=0,87 (p<0,001), CCI=0,96 (IC 0,94–0,97). El análisis de Bland-Altman mostró una media de las diferencias de−0,9mmHg (IC−2,0 a 0,2mmHg). La correlación fue mejor en ausencia de patología respiratoria, con asistencia respiratoria baja, con PaCO2>50mmHg y con aplicación frontal del sensor. Hubo consistencia de los resultados. No se observaron efectos secundarios derivados de la utilización del anillo.

Conclusiones:

La correlación obtenida entre la PtCO2 y PaCO2 fue muy buena. El monitor digital SenTec y el sensor de anillo específico constituyen una herramienta fiable, segura y fácil de utiliza (AU)
ABSTRACT

Objective:

To estimate the accuracy of the transcutaneous carbon dioxide tension measurement (PtCO2) compared to the measurement of the arterial carbon dioxide tension (PaCO2). Material and

methods:

An analytical, longitudinal, prospective and observational study, of a dynamic cohort taken from the in-patients of a Paediatric Intensive Care Unit (PICU). The PtCO2 was measured with the SenTec AG analyzer, and the sensor was applied with the specific Multi-Site Attachment Ring. PtCO2 and PaCO2 were recorded at the same time. The statistical significance of the association between paired measurements was evaluated with the Snedecor's F test, the Pearson's r2 correlation coefficient and the Interclass Correlation Coefficient (ICC). The degree of agreement was evaluated with the Bland & Altman method. The consistency of the results was evaluated with the ANalysis Of the VAriance (ANOVA).

Results:

One hundred and six paired measurements, PtCO2 and PaCO2, from twelve patients, were compared. The means of the PaCO2 and PtCO2 were 51.0±13mmHg and 50.1±14mmHg, respectively; r2=0.87 (p<0.001), ICC=0.96, (95% CI 0.94–0.97). The Bland-Altman analysis showed a mean difference of−0.9mmHg (95% CI−2.0 to 0.2mmHg). The correlation was better in cases with no respiratory disease, with low respiratory assistance, with PaCO2>50mmHg and with the sensor applied on the forehead. The results were consistent. No side effects derived from the use of the ring were observed.

Conclusion:

The correlation between PtCO2 and PaCO2 is excellent and stable. The ring sensor was safe and easy to us (AU)
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Pressão Parcial / Dióxido de Carbono / Cuidados Críticos / Monitorização Fisiológica Tipo de estudo: Estudo observacional / Fatores de risco Limite: Criança / Feminino / Humanos / Masculino Idioma: Espanhol Revista: An. pediatr. (2003, Ed. impr.) Ano de publicação: 2010 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario de Salamanca/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Pressão Parcial / Dióxido de Carbono / Cuidados Críticos / Monitorização Fisiológica Tipo de estudo: Estudo observacional / Fatores de risco Limite: Criança / Feminino / Humanos / Masculino Idioma: Espanhol Revista: An. pediatr. (2003, Ed. impr.) Ano de publicação: 2010 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario de Salamanca/España
...