Your browser doesn't support javascript.
loading
Complications associated with the ball, bar and Locator attachmentsfor implant-supported overdentures
Cakarer, Sirmahan; Can, Taylan; Yaltirik, Mehmet; Keskin, Cengizhan.
Afiliação
  • Cakarer, Sirmahan; Istanbul University. Dentistry Faculty. Department of Oral and Maxillofacial Surgery. s. c. s. p
  • Can, Taylan; Istanbul University. Dentistry Faculty. Department of Oral and Maxillofacial Surgery. s. c. s. p
  • Yaltirik, Mehmet; Istanbul University. Dentistry Faculty. Department of Oral and Maxillofacial Surgery. s. c. s. p
  • Keskin, Cengizhan; Istanbul University. Dentistry Faculty. Department of Oral and Maxillofacial Surgery. s. c. s. p
Med. oral patol. oral cir. bucal (Internet) ; 16(7): 953-959, .nov. 2011. tab
Artigo em Inglês | IBECS | ID: ibc-93494
Biblioteca responsável: ES1.1
Localização: BNCS
ABSTRACT

Background:

The purpose of this clinical study was to evaluate the complications associated with the different attachments used in implant-supported overdentures, including prosthetic problems and implant failures. A comparison of ball, bar and Locator (Zest Anchors, Inc, homepage, Escondido, CA, USA) attachments, in completely edentulous patients with two, three or four implants, was conducted. Material and

methods:

A total of 36 edentulous patients (20 female, 16 male) with a mean age of 66.3 years, were enrolled in the study. The patients were treated with 95 implants, for the prosthetic restoration of the maxilla or the mandible. The mean follow-up time was 41.17 months. Prosthetic complications including, fractured overdentures,replacements of O-ring attachment and retention clips, implant failures, hygiene problems, mucosal enlargements, attachment fractures, retention loss and dislodgement of the attachments were recorded and evaluated.The recall visits at 3, 6, 12 months and, annually there after.

Results:

Fourteen complications in the ball attachment group and 7 complications in the bar group were observed.No complications were observed in the locator group. The difference was found to be as statistically significant(p=0,009). Six of the 95 implants had failed. Totally 39 implant overdentures were applied. Three prostheses wererenewed because of fractures.

Conclusion:

Within the limits of the present study, it was concluded that the locator system showed superior clinical results than the ball and the bar attachments, with regard to the rate of prosthodontic complications and the maintenance of the oral function (AU)
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Coroas / Implantação Dentária Tipo de estudo: Fatores de risco Limite: Humanos Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2011 Tipo de documento: Artigo Instituição/País de afiliação: Istanbul University/s. p
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Coroas / Implantação Dentária Tipo de estudo: Fatores de risco Limite: Humanos Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2011 Tipo de documento: Artigo Instituição/País de afiliação: Istanbul University/s. p
...