Your browser doesn't support javascript.
loading
Stents auto-expansíveis versus balão-expansível para tratamento de lesões em vasos de fino calibre: estudo com ultra-som intracoronário tridimensional / Self-expanding versus balloon-expandable stents for the treatment of small coronary vessels: a three-dimensional intracoronary ultrasound study
Chamié, Daniel; Costa Junior, José Ribamar; Abizaid, Alexandre; Feres, Fausto; Devito, Fernando; Mattos, Luiz Alberto; Staico, Rodolfo; Abizaid, Andrea; Tanajura, Luiz Fernando L; Fitzgerald, Peter J; Whitbourn, Robert; Sousa, Amanda G. M. R; Sousa, J. Eduardo.
Afiliação
  • Chamié, Daniel; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Costa Junior, José Ribamar; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Abizaid, Alexandre; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Feres, Fausto; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Devito, Fernando; Hospital São Domingos. Faculdade de Medicina da FIPA. Catanduva. BR
  • Mattos, Luiz Alberto; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Staico, Rodolfo; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Abizaid, Andrea; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Tanajura, Luiz Fernando L; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Fitzgerald, Peter J; Stanford University Medical Center. California. US
  • Whitbourn, Robert; St Vicent Hospital.
  • Sousa, Amanda G. M. R; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Sousa, J. Eduardo; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
Rev. bras. cardiol. invasiva ; 16(3): 341-352, jul.-set. 2008. ilus, graf, tab
Artigo em Português | LILACS, Sec. Est. Saúde SP | ID: lil-503483
Biblioteca responsável: BR44.1
RESUMO

Introdução:

Intervenção coronária pecutânea em vasos de fino calibre (VF) está associada a piores resultados imediatos e tardios, com elevadas taxas de reestenose. Estudos prévios têm sugerido que stents auto-expansíveis causam menos injúria vascular no momento do implante, com expansão de seus volumes com o tempo, gerando maiores áreas luminais que stents balão-expansíveis. A influência desses fenômenos em VF ainda é desconhecida.

Objetivos:

Avaliar as propriedades mecânicas e a eficácia do novo stent CardioMind no tratamento de lesões em VF em comparação ao stent balão-expansível Multi-link Pixel (Pixel).

Método:

Treze pacientes portadores de lesões únicas primárias < 14 mm de extensão, em artérias coronárias nativas < 2,5 mm de diâmetro, foram tratados com o stent CardioMind e comparados a uma corte histórica de 25 pacientes, com os mesmos critérios de inclusão, tratados com o stent Pixel. Ultrasom intracoronário (USIC) seriado foi realizado pós-procedimento e aos 7,3 +/- 1,0 meses.

Resultados:

A média das idades foi de 58,1 +/- 9,9 anos, com 60,5 por cento do sexo masculino e 39,4 por cento diabéticos. Ambos os stents produziram volumes de hiperplasia neo-intimal (HNI) smelhantes...
ABSTRACT

Background:

Percutaneous coronary intervention in small vessels (SV) is associated with poor short- and long-term outcomes, with high rates of restenosis. Previous studies have suggested that self-expanding stents can cause less vessel injury at implantation, expanding their volumes over time, and leading to larger luminal areas than those of balloon-expandable stents. The influence of these phenomena on SV remains unknown.

Objectives:

To assess the mechanical properties and efficacy of the novel CardioMind™ stent in comparison with the balloon-expandable Multi-Link Pixel™ (Pixel) stent in the treatment of SV.

Methods:

Thirteen patients with single, de novo, < 14 mm length lesions in native coronary arteries < 2.5 mm in diameter were treated with the CardioMind™ stent and compared with a historical cohort of 25 patients, with the same inclusion criteria, treated with the Pixel™ stent. Intravascular ultrasound (IVUS) was performed serially after the procedure and at 7.3 ± 1.0 months of follow-up.

Results:

Mean age was 58.1 ± 9.9 years; 60.5% were male and 39.4% were diabetic. Both stents produced similar neointimal hyperplasia (NIH) volumes (indexed NIH volume 1.45 ± 0.46 mm³/mm for CardioMind™ versus 1.66 ± 1.02 mm³/mm for Pixel™; p = 0.48). However, the CardioMind™ stentpresented a 12% expansion of its volume, leading to a...
Assuntos
Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: LILACS / Sec. Est. Saúde SP Assunto principal: Angioplastia Coronária com Balão / Stents Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. cardiol. invasiva Ano de publicação: 2008 Tipo de documento: Artigo Instituição/País de afiliação: Hospital São Domingos/BR / Instituto Dante Pazzanese de Cardiologia/BR / Stanford University Medical Center/US
Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: LILACS / Sec. Est. Saúde SP Assunto principal: Angioplastia Coronária com Balão / Stents Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. cardiol. invasiva Ano de publicação: 2008 Tipo de documento: Artigo Instituição/País de afiliação: Hospital São Domingos/BR / Instituto Dante Pazzanese de Cardiologia/BR / Stanford University Medical Center/US
...