Evaluation of the automated system Vitek2 for identification and antimicrobial susceptibility testing of Brazilian Gram-positive cocci strains
d'Azevedo, Pedro Alves; Siquiera, Itacy; Gugel, Juliana; Antunes, Ana Lúcia Souza; Secchi, Carina; Pasternak, Jacyr; Martino, Marinês Dalla Vale.
Afiliação
  • d'Azevedo, Pedro Alves; Foundation of Clinical Medical School of Porto Alegre. Department of Microbiology and Parasitology. Porto Alegre. BR
  • Siquiera, Itacy; Hospital Israelita Albert Einstein. Clinical Laboratory. São Paulo. BR
  • Gugel, Juliana; Hospital Israelita Albert Einstein. Clinical Laboratory. São Paulo. BR
  • Antunes, Ana Lúcia Souza; Foundation of Clinical Medical School of Porto Alegre. Department of Microbiology and Parasitology. Porto Alegre. BR
  • Secchi, Carina; Foundation of Clinical Medical School of Porto Alegre. Department of Microbiology and Parasitology. Porto Alegre. BR
  • Pasternak, Jacyr; Hospital Israelita Albert Einstein. Clinical Laboratory. São Paulo. BR
  • Martino, Marinês Dalla Vale; Hospital Israelita Albert Einstein. Clinical Laboratory. São Paulo. BR
Braz. j. infect. dis ; 13(2): 107-110, Apr. 2009. tab
Article em En | LILACS | ID: lil-538214
Biblioteca responsável: BR1.1
ABSTRACT
Automated instruments offer many advantages for clinical laboratories. Nevertheless, they can have problems identifying and determining susceptibilities of some pathogens. Vitek® 2 (bioMérieux) is an automated system that was recently introduced to Brazil. We evaluated the performance of this equipment for Brazilian isolates that had been characterized using reference identification and antimicrobial susceptibility testing methods. Ninety-nine strains of Gram-positive cocci from a local reference center collection were analyzed, consisting of 50 coagulasenegative Staphylococcus (CoNS) and 49 Enterococcus and related species. Vitek® 2 correctly identified 79.8 percent (79/99) of the isolates. Oxacillin resistance was detected in 76 percent (19/25) of resistant S. epidermidis strains and in 88 percent (22/25) of other resistant CoNS species strains. Vancomycin resistance was detected in 100 percent (20/20) of resistant Enterococcus and related species strains. Vitek® 2 performed very well for the identification of S. epidermidis and non-epidermidis staphylococci, and for the detection of vancomycin resistance in Enterococcus and related species. However, the system needs improvement in order to provide reliable results for the characterization of some CoNS species, identification of Enterococcus and related species and for detecting oxacillin resistance in CoNS.
Assuntos
Palavras-chave
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Staphylococcus / Enterococcus / Automação Laboratorial / Antibacterianos Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans País/Região como assunto: America do sul / Brasil Idioma: En Revista: Braz. j. infect. dis Assunto da revista: DOENCAS TRANSMISSIVEIS Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Brasil
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Staphylococcus / Enterococcus / Automação Laboratorial / Antibacterianos Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans País/Região como assunto: America do sul / Brasil Idioma: En Revista: Braz. j. infect. dis Assunto da revista: DOENCAS TRANSMISSIVEIS Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Brasil