Your browser doesn't support javascript.
loading
Caracterização dosimétrica de feixes de fótons com detector de diamante / Dosimetric characterization of photon beams using a diamond detector
Sabino, Talita; Rodrigues, Laura Natal; Furnari, Laura; Watanabe, Érika Yumi; Menegussi, Gisela.
Afiliação
  • Sabino, Talita; s.af
  • Rodrigues, Laura Natal; s.af
  • Furnari, Laura; s.af
  • Watanabe, Érika Yumi; s.af
  • Menegussi, Gisela; s.af
Radiol. bras ; Radiol. bras;45(1): 35-38, jan.-fev. 2012. ilus, graf, tab
Article em Pt | LILACS | ID: lil-618393
Biblioteca responsável: BR734.1
RESUMO

OBJETIVO:

Este trabalho foi realizado com o objetivo de caracterizar o detector de diamante, bem como observar seu comportamento.

MATERIAIS E MÉTODOS:

As características dosimétricas de feixes de fótons de 6 MV de energia foram medidas utilizando o sistema automático de varredura MP3 da PTW com um detector de diamante e foram comparadas com medidas usando uma câmara de ionização 31010 da PTW.

RESULTADOS:

As características dosimétricas do detector de diamante foram observadas por meio de medidas de linearidade com a dose, dependência com a taxa de dose e distribuições de dose em profundidade, bem como perfis.

CONCLUSÃO:

Algumas medidas com diamante e câmara de ionização foram validadas com resultados publicados na literatura, o que demonstrou um bom comportamento do detector de diamante na comparação com a câmara de ionização, muito utilizada para dosimetria em radioterapia, evidenciando que o diamante é uma boa escolha de detector para dosimetria de campos pequenos.
ABSTRACT

OBJECTIVE:

The present study was aimed at characterizing a diamond detector and observing its behavior. MATERIALS AND

METHODS:

The dosimetric characteristics of 6 MV photon beams were measured by utilizing an automatic PTW MP3 scanning system with a diamond detector, and compared with measurements performed with a PTW 31010 ion chamber.

RESULTS:

Measurements of dose linearity, dose rate dependence and depth dose distribution as well as dose profiles demonstrated the dosimetric characteristics of the diamond detector.

CONCLUSION:

Some measurements with diamond detector and ion chamber were validated with results published in the literature, demonstrating a good behavior of the diamond detector as compared with the ionization chamber that is widely utilized for dosimetry in radiotherapy, indicating that the diamond detector is a good choice for small field dosimetry.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Radioterapia / Diamante / Dosimetria / Técnicas e Procedimentos Diagnósticos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: Pt Revista: Radiol. bras Assunto da revista: DIAGNOSTICO POR IMAGEM / RADIOLOGIA Ano de publicação: 2012 Tipo de documento: Article País de publicação: Brasil

Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Radioterapia / Diamante / Dosimetria / Técnicas e Procedimentos Diagnósticos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: Pt Revista: Radiol. bras Assunto da revista: DIAGNOSTICO POR IMAGEM / RADIOLOGIA Ano de publicação: 2012 Tipo de documento: Article País de publicação: Brasil