Your browser doesn't support javascript.
loading
Complicaciones del implante de marcapaso definitivo: ¿Un evento operador dependiente? Análisis de 743 pacientes consecutivos / Complications of permanent pacemaker: Event operator dependent? Analysis of 743 consecutive patients
Femenía, Francisco J.; Arce, Mauricio; Peñafort, Fernando; Arrieta, Martín; Gutiérrez, Daniel.
Afiliação
  • Femenía, Francisco J.; Hospital Español de Mendoza. Departamento de Cardiología. Unidad de Arritmias. AR
  • Arce, Mauricio; Hospital Español de Mendoza. Departamento de Cardiología. Unidad de Arritmias. AR
  • Peñafort, Fernando; Hospital Español de Mendoza. Departamento de Cardiología. Unidad de Arritmias. AR
  • Arrieta, Martín; Hospital Español de Mendoza. Departamento de Cardiología. Unidad de Arritmias. AR
  • Gutiérrez, Daniel; Hospital Español de Mendoza. Departamento de Cardiología. Unidad de Arritmias. AR
Arch. cardiol. Méx ; Arch. cardiol. Méx;80(2): 95-99, abr.-jun. 2010. tab
Article em Es | LILACS | ID: lil-631966
Biblioteca responsável: MX1.1
RESUMEN
La frecuencia de complicaciones relacionadas al implante de marcapasos, es variable y se encuentra entre 6% a 9 %, y existen diferentes factores relacionados con las mismas.

Objetivos:

Evaluar las distintas complicaciones relacionadas con el implante de marcapasos y determinar cuáles son los factores de riesgo independientes que las predisponen.

Métodos:

Análisis descriptivo, tipo casos-controles, retrospectivo y prospectivo, correspondiente a 743 consecutivos, con implante de marcapasos definitivo, entre enero de 2007 y abril de 2008.

Resultados:

Edad de 73.83 ± 12 años, 63.9% varones. Los pacientes con algún grado de cardiopatía estructural representaron 43%. No hubo muertes relacionadas al implante de marcapasos. Tuvieron complicaciones relacionadas con el implante 57 pacientes (7.67%); 34 casos (4.57%), requirieron de una re-intervención quirúrgica. Hubo complicaciones graves un paciente presentó endocarditis infecciosa (0.13%) que motivó la extracción del dispositivo y otro paciente, luego de la punción venosa subclavia, presentó neumotórax con necesidad de drenaje pleural (0.13%). En el análisis multivariado, las variables independientes relacionadas fueron un operador poco experimentado y la presencia de insuficiencia cardiaca previa al implante.

Conclusiones:

En esta serie consecutiva de pacientes con implante de marcapasos definitivo, la incidencia de complicaciones se correlacionó con 1) operador con poca experiencia y 2) la presencia de insuficiencia cardiaca previa al implante, se relacionaron con las mismas.
ABSTRACT
The complications of permanent pacemaker implant, are found among 6%-9% of patients. Different factors are related.

Objectives:

We evaluated the complications related with pacemaker implants and which are the independent risk factors associated with them.

Methods:

Descriptive analysis, type case-controls, retrospective and prospective, including 743 consecutive patients, between January 2007 and April 2008.

Results:

Age 73.83 ± 12 years, 63.9% male. 43% of the patients had some degree of structural cardiopathy. There were no deaths related to pacemaker implants. Implant procedure-related complications were observed in 57 patients (7.67%), requiring a surgical intervention in 34 patients (4.57%). The most serious complications included one device-related infective endocarditis (0.13%), the entire device system was extracted, and one pneumothorax following subclavian vein puncture with insertion of a chest tube. In the multivariate analysis, low level of operator experience and preimplant heart failure were independent predictors for complications.

Conclusions:

In this consecutive series of patients with permanent pacemaker implant, there was a low rate of complications, similar to the reported international series. A low level of operator experience and preimplant heart failure were related.
Assuntos
Palavras-chave
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Marca-Passo Artificial Tipo de estudo: Etiology_studies / Observational_studies / Prevalence_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: Es Revista: Arch. cardiol. Méx Assunto da revista: CARDIOLOGIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Argentina País de publicação: México
Texto completo: 1 Coleções: 01-internacional Base de dados: LILACS Assunto principal: Marca-Passo Artificial Tipo de estudo: Etiology_studies / Observational_studies / Prevalence_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male Idioma: Es Revista: Arch. cardiol. Méx Assunto da revista: CARDIOLOGIA Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Argentina País de publicação: México