Estudio comparativo en pacientes usuarios de implantes Vibrant Soundbridge® (VSB) y sistema osteointegrado BAHA® / Comparative study on patients users of Vibrant Soundbridge® and osseointegrated BAHA® implants
Rivas, José Antonio; Velandia, Román; Rincón, Luz Adriana; Tamayo, César Augusto; Forero, Víctor Hugo; Rivas, Adriana; Rivas, Alejandro.
Afiliação
  • Rivas, José Antonio; Clínica José A. Rivas. CO
  • Velandia, Román; Clínica José A. Rivas. CO
  • Rincón, Luz Adriana; Clínica José A. Rivas. CO
  • Tamayo, César Augusto; Clínica José A. Rivas. CO
  • Forero, Víctor Hugo; Clínica José A. Rivas. CO
  • Rivas, Adriana; Clínica José José A. Rivas. CO
  • Rivas, Alejandro; Vanderbilt University. Nashville. US
Acta otorrinolaringol. cir. cabeza cuello ; 40(3): 194-201, jul.-sept. 2012. ilus
Artigo em Espanhol | LILACS | ID: lil-683635
Biblioteca responsável: CO361.9
RESUMEN

Objetivo:

presentar la experiencia de una institución en la rehabilitación de pacientes con pérdida auditiva utilizando el implante de oído medio Vibrant Soundbridge® (VSB).

Diseño:

estudio retrospectivo. Materiales y

métodos:

se revisaron registros médicos, quirúrgicos y audiológicos de pacientes a los que se les implantó el sistema Vibrant Soundbridge (VSB), evaluando la ganancia auditiva establecida por la diferencia en los promedios de las frecuencias 0.5 a 3 kHz en la audiometría tonal postimplante con el sistema y sin él. Se evaluó también la satisfacción de uso con el dispositivo mediante la encuesta Escala de Satisfacción de Audición con Dispositivo (HDSS, Hearing Device Satisfaction Scale). Realizamos comparación de desenlaces audiológicos con una población de usuarios de similares características del sistema osteointegrado BAHA®.

Resultados:

25 pacientes con pérdida auditiva predominantemente mixta recibieron un implante VSB colocando el Transductor de Masa Flotante (FMT) en ventana redonda en 23 casos (92%). La ganancia auditiva promedio fue de 36 ± 16 decibeles (dB). No se identificaron diferencias significativas en la ganancia auditiva entre los procesadores AP404-36 dB y Amadé-38 dB, p = 0.6. La ganancia auditiva en el grupo de usuarios del sistema BAHA fue de 37 dB y la discriminación de lenguaje a 65 dB fue de 93% y 100% para VSB y BAHA® respectivamente. En ambos sistemas el puntaje global de satisfacción fue de 4/5.
ABSTRACT

Objective:

Present the institutional experience on the auditory rehabilitation of patients implanted with the middle ear prosthesis Vibrant Soundbridge®.

Design:

Retrospective study of a cases series. Materials and

Methods:

Medical, surgical and audiological charts review of patients implanted with the middle ear system VSB. Hearing gain was established by the mean differences calculated on 0.5 to 3 kHz in the pure tone audiometries pre and post implantation, with and without the device. Satisfaction with the device was assessed with the Hearing Device Satisfaction Scale tool. In addition, audiological and satisfaction comparison with a population with similar characteristics of BAHA implant users was made.

Results:

25 patients with conductive and mixed hearing loss were implanted with the VSB, placing the FMT on the round window in 23cases (92%). Mean auditory gain was 36 ± 16 decibels (dB). There were no significance gains between processors AP404-36 dB and Amadé-38 dB, p = 0.6. Mean auditory gain in the BAHA® users was 37 dB; speech discrimination at 65 dB was 93% and 100% for VSB and BAHA® users respectively. Overall satisfaction score for both systems was 4/5.

Conclusion:

The middle ear implant system VSB gives a satisfactory hearing gain to patients with moderate mixed hearing loss comparable with the BAHA® as well in satisfaction and quality of life issues.
Assuntos

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Prótese Ossicular / Surdez / Implantes Auditivos de Tronco Encefálico Tipo de estudo: Estudo observacional / Estudo prognóstico Aspecto: Preferência do paciente Limite: Humanos Idioma: Espanhol Revista: Acta otorrinolaringol. cir. cabeza cuello Assunto da revista: Lesäes do Pesco‡o / Otolaringologia / Otorrinolaringopatias / Traumatismos Craniocerebrais Ano de publicação: 2012 Tipo de documento: Artigo País de afiliação: Colômbia / Estados Unidos Instituição/País de afiliação: Clínica José A. Rivas/CO / Clínica José José A. Rivas/CO / Vanderbilt University/US
Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Prótese Ossicular / Surdez / Implantes Auditivos de Tronco Encefálico Tipo de estudo: Estudo observacional / Estudo prognóstico Aspecto: Preferência do paciente Limite: Humanos Idioma: Espanhol Revista: Acta otorrinolaringol. cir. cabeza cuello Assunto da revista: Lesäes do Pesco‡o / Otolaringologia / Otorrinolaringopatias / Traumatismos Craniocerebrais Ano de publicação: 2012 Tipo de documento: Artigo País de afiliação: Colômbia / Estados Unidos Instituição/País de afiliação: Clínica José A. Rivas/CO / Clínica José José A. Rivas/CO / Vanderbilt University/US
...