Rapid method for the determination of trace fluoride and activation of ion-selective electrode.
Anal Sci
; 19(5): 671-3, 2003 May.
Article
em En
| MEDLINE
| ID: mdl-12769362
A method for activating ISE is proposed that can allow determination of the fluoride concentration at ng mL(-1) level with good precision and accuracy. Fluoride ISE is activated in 0.5 mol L(-1) HClO4 medium and then fluoride is determined in the same medium. The linear range for the determination of fluoride is between 1.00 x 10(-2)-1.00 x 10(-7) mol L(-1), and the detection limit of the method is 1.0 ng mL(-1). The advantage of this method is that it is free from the use of TISAB solution while being, time-saving and labor-saving. A mechanism study of the activation of FISE in HClO4 medium is explained. The method has been used for the determination of trace fluoride in milk and flour with satisfactory results.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
Anal Sci
Ano de publicação:
2003
Tipo de documento:
Article
País de afiliação:
China
País de publicação:
Suíça