The accuracy of dynamic wedge dose computation in the ADAC Pinnacle RTP system.
Shao, H; Wu, X; Luo, C; Crooks, A; Bernstein, A; Markoe, A.
Afiliação
  • Shao H; Radiation Oncology Department, University of Miami, School of Medicine, 1475 N.W. 12th Avenue, Miami, Florida 33136, USA. huashao@hotmail.com
J Appl Clin Med Phys ; 5(4): 46-54, 2004.
Article em En | MEDLINE | ID: mdl-15738920
The nonphysical wedge is a modality that uses computer-controlled jaw motion to generate wedge-shaped dose distributions. There are Varian enhanced dynamic wedges (EDWs) and Siemens virtual wedges (VWs). We recently commissioned dynamic wedges on both Varian and Siemens LINACs. The beam data, acquired with a Wellhofer chamber array and a Sun Nuclear profiler, are used for modeling in the ADAC Pinnacle system. As recommended by ADAC, only a limited number of beam data is measured and used for beam modeling. Therefore, the dose distributions of dynamic wedges generated by Pinnacle must be examined. Following the commissioning of the dynamic wedges, we used Pinnacle to generate a number of dose distributions with different energies, wedge angles, field sizes, and depths. The computed data from Pinnacle are then compared with the measured data. The deviations of the output factor in all square and rectangular fields are mostly within 2.0% for both EDW and VW. For asymmetric fields, the deviations are within 3%. However, exceptions of differences more than 3% have been found in a larger field and large wedge combinations. The precision of the beam profiles generated by Pinnacle is also evaluated. As a result of this investigation, we present a scope of quality assurance tests that are necessary to ensure acceptable consistency between the delivered dose and the associated treatment plan when dynamic wedges are applied.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Radioterapia Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Radioterapia Conformacional Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Radioterapia Assistida por Computador / Planejamento da Radioterapia Assistida por Computador / Radioterapia Conformacional Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos