Influence of photoactivation protocol and light guide distance on conversion and microleakage of composite restorations.
Oper Dent
; 34(4): 408-14, 2009.
Article
em En
| MEDLINE
| ID: mdl-19678445
PURPOSE: To evaluate the effect of light guide distance and the different photoactivation methods on the degree of conversion (DC) and microleakage of a composite. METHODS AND MATERIALS: Three photoactivation protocols (600 mW/cm2 x 40 seconds; 400 mW/cm2 x 60 seconds or 200 mW/cm2 x 20 seconds, followed by 500 mW/cm2 x 40 seconds) and three distances from the light source (0, 3 or 7 mm) were tested. Cylindrical specimens (5 mm diameter; 2 mm tall; n=3) were prepared for the DC test (FT-Raman). Class V cavities were made in 90 bovine incisors to conduct the microleakage test. The specimens were conditioned for 15 seconds with phosphoric acid (37%), followed by application of the adhesive system Prime & Bond NT (Dentsply/Caulk). The preparations were restored in bulk. The specimens were stored for 24 hours in distilled water (37 degrees C) before being submitted to the silver-nitrate microleakage protocol. The restorations were sectioned and analyzed under 25x magnification. RESULTS: Statistical analyses (two-way ANOVAs and Tukey test, alpha=0.05) found significance only for the factor distance (p=0.015) at the top of the composite for the DC test. Conversion was statistically lower for the 7 mm groups compared to the 0 and 3 mm groups, which were equivalent to each other. At the bottom of the specimens, none of the factors or interactions was significant (p<0.05). The Kruskal-Wallis test showed that, in general, the soft-start method led to lower microleakage scores when compared to the continuous modes, mainly when associated with a distancing of 7 mm (p<0.01). With the exception of specimens irradiated with 400 mW/cm2 that did not demonstrate variations on scores for the distances tested, higher microleakage was observed for shorter distances from the light source. CONCLUSIONS: Soft-start methods may reduce microleakage when the light guide distancing provides a low level of irradiance, which also causes a discrete reduction in the DC.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Poliuretanos
/
Resinas Acrílicas
/
Resinas Compostas
/
Infiltração Dentária
/
Restauração Dentária Permanente
/
Cura Luminosa de Adesivos Dentários
Limite:
Animals
Idioma:
En
Revista:
Oper Dent
Ano de publicação:
2009
Tipo de documento:
Article
País de afiliação:
Brasil
País de publicação:
Estados Unidos