Your browser doesn't support javascript.
loading
Multiple double cross-section transmission electron microscope sample preparation of specific sub-10 nm diameter Si nanowire devices.
Gignac, Lynne M; Mittal, Surbhi; Bangsaruntip, Sarunya; Cohen, Guy M; Sleight, Jeffrey W.
Afiliação
  • Gignac LM; IBM Thomas J. Watson Research Center, 1101 Kitchawan Road, Yorktown Heights, NY 10598, USA. gignac@us.ibm.com
Microsc Microanal ; 17(6): 889-95, 2011 Dec.
Article em En | MEDLINE | ID: mdl-22071222
The ability to prepare multiple cross-section transmission electron microscope (XTEM) samples from one XTEM sample of specific sub-10 nm features was demonstrated. Sub-10 nm diameter Si nanowire (NW) devices were initially cross-sectioned using a dual-beam focused ion beam system in a direction running parallel to the device channel. From this XTEM sample, both low- and high-resolution transmission electron microscope (TEM) images were obtained from six separate, specific site Si NW devices. The XTEM sample was then re-sectioned in four separate locations in a direction perpendicular to the device channel: 90° from the original XTEM sample direction. Three of the four XTEM samples were successfully sectioned in the gate region of the device. From these three samples, low- and high-resolution TEM images of the Si NW were taken and measurements of the NW diameters were obtained. This technique demonstrated the ability to obtain high-resolution TEM images in directions 90° from one another of multiple, specific sub-10 nm features that were spaced 1.1 µm apart.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Silício / Manejo de Espécimes / Nanotecnologia / Nanofios / Micromanipulação Idioma: En Revista: Microsc Microanal Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Silício / Manejo de Espécimes / Nanotecnologia / Nanofios / Micromanipulação Idioma: En Revista: Microsc Microanal Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Reino Unido