Development of a FI-HG-ICP-OES solid phase preconcentration system for inorganic selenium speciation in Argentinean beverages.
Food Chem
; 169: 73-9, 2015 Feb 15.
Article
em En
| MEDLINE
| ID: mdl-25236200
A preconcentration system has been developed to determine inorganic selenium species. Selenium was retained by a column filled with polyvinyl chloride (PVC) with lanthanum hydroxide co-precipitation. Speciation was achieved by selective photoreduction previous Se preconcentration. The retention pH was optimized at 10.0. Two multivariate calibrations and a central composite design were employed for optimization of the system. Sample, reagents and acid flow rates are significant variables affecting the system. Employing HG-ICP-OES as detection, the optimized system reached a detection limit of 0.03µg/L, and an enhancement factor of 14875 (25 for preconcentration system, 595 for hydride generation). To verify the method' accuracy, two certified reference materials, BCR® 414 Plankton & IRMM-804 Rice Flour, were analysed. The system was applied to inorganic selenium speciation in several Argentinean beverages to estimate their selenium contribution to diet.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Espectrometria de Massas
/
Bebidas
/
Compostos de Selênio
/
Extração em Fase Sólida
Tipo de estudo:
Evaluation_studies
País/Região como assunto:
America do sul
/
Argentina
Idioma:
En
Revista:
Food Chem
Ano de publicação:
2015
Tipo de documento:
Article
País de afiliação:
Argentina
País de publicação:
Reino Unido