Reference drums used in calibration of a plastic scintillation counter in a 4π counting geometry.
Appl Radiat Isot
; 109: 105-108, 2016 Mar.
Article
em En
| MEDLINE
| ID: mdl-26651167
In this study, two kinds of reference drums were developed. One type was constructed with nine layers of large-area sources filled with different materials having five different densities. The other type of reference drums was constructed with nine rod sources filled with the same materials of different densities. The efficiency calibration of a plastic scintillation counter in 4π counting geometry using these two kinds of drums showed that rod-source drums resulted in higher counting efficiency than layered source drums. The counting rates obtained from rod-source drums were closer to those obtained from a standard drum with water solution than counting rates from drums with layered sources. The results of this study recommend to use reference drums with rod-sources to compensate the drawbacks of standard drums with water solution of not being able to adjust the density of material. The proposed reference drums improve the accuracy of radioactivity analysis for waste drums of different densities.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Tipo de estudo:
Prognostic_studies
Idioma:
En
Revista:
Appl Radiat Isot
Assunto da revista:
MEDICINA NUCLEAR
/
SAUDE AMBIENTAL
Ano de publicação:
2016
Tipo de documento:
Article
País de publicação:
Reino Unido