Determination of elemental impurities in poly(vinyl chloride) by inductively coupled plasma optical emission spectrometry.
Pereira, Leticia S F; Pedrotti, Matheus F; Miceli, Tatiane M; Pereira, Juliana S F; Flores, Erico M M.
Afiliação
  • Pereira LS; Departamento de Química, Universidade Federal de Santa Maria, 97105-900 Santa Maria, RS, Brazil.
  • Pedrotti MF; Departamento de Química, Universidade Federal de Santa Maria, 97105-900 Santa Maria, RS, Brazil.
  • Miceli TM; Departamento de Química Inorgânica, Universidade Federal do Rio Grande do Sul, 91501-970 Porto Alegre, RS, Brazil.
  • Pereira JS; Departamento de Química Inorgânica, Universidade Federal do Rio Grande do Sul, 91501-970 Porto Alegre, RS, Brazil.
  • Flores EM; Departamento de Química, Universidade Federal de Santa Maria, 97105-900 Santa Maria, RS, Brazil. Electronic address: ericommf@gmail.com.
Talanta ; 152: 371-7, 2016 May 15.
Article em En | MEDLINE | ID: mdl-26992533
In this work, a method for poly(vinyl chloride) (PVC) analysis by inductively coupled plasma optical emission spectrometry (ICP-OES) was developed. Samples were digested by microwave-induced combustion (MIC) and thirteen elements (Ba, Cd, Co, Cu, Fe, Mg, Mn, Mo, Ni, Pb, Sr, V and Zn) were determined by ICP-OES. Operational conditions of MIC were investigated allowing quantitative determination of all the analytes and suitable results were achieved using a 3 mol L(-1) HNO3 solution. Microwave-assisted wet digestion (MAWD) using closed quartz vessels and a single reaction chamber microwave digestion system (MAWD-SRC), were also evaluated for PVC digestion for results comparison. All the evaluated sample preparation methods were considered suitable for PVC digestion but MIC was preferable due to the possibility of using diluted HNO3 instead of concentrated reagents and due to the better digestion efficiency. The residual carbon content (RCC) in digests obtained by MIC was significantly lower in comparison with the values obtained after MAWD and by MAWD-SRC. Accuracy for the proposed method was better than 94% for all analytes by comparison of results with those obtained by neutron activation analysis (NAA). Using MIC, it was possible to digest a relatively high sample mass (500 mg) and up to 8 samples in less time (25 min) in comparison with MAWD and MAWD-SRC (about 60 min for both methods). The efficiency of digestion by MIC was higher 99% and lower limits of detection (as low as 0.1 µg g(-1)) were obtained avoiding the use of concentrated acids that is of great concern according to the current green chemistry recommendations.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cloreto de Polivinila / Análise Espectral / Fenômenos Ópticos / Gases em Plasma Idioma: En Revista: Talanta Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Brasil País de publicação: Holanda