QAMaster: A new software framework for phantom-based computed tomography quality assurance.
Karius, Andre; Bert, Christoph.
Afiliação
  • Karius A; Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Bert C; Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Erlangen, Germany.
J Appl Clin Med Phys ; 23(4): e13588, 2022 Apr.
Article em En | MEDLINE | ID: mdl-35297569
The regular evaluation of imaging performance of computed tomography (CT) scanners is essential for CT quality assurance. For automation of this process, the software QAMaster was developed at the Universitätsklinikum Erlangen, which provides based on CT scans of the CatPhan® 504 (The Phantom Laboratory, Salem, USA) automated image quality analysis and documentation by evaluating CT number accuracy, spatial linearity, uniformity, contrast-noise-ratio, spatial resolution, noise, and slice thickness. Dose assessment is supported by calculations of the weighted computed tomography dose index (CTDIw ) and weighted cone beam dose index (CBDIw ). QAMaster was tested with CatPhan® 504 scans and compared to manual evaluations of these scans, whereby high consistency of the respective results was observed. The CT numbers, spatial linearity, uniformity, contrast-noise-ratio, noise, and slice thickness deviated by only (0.13 ± 0.25) HU, (0.02 ± 0.05) mm, (-0.01 ± 0.03)%, 0.8 ± 1.8, (0.131 ± 0.05) HU, and (0.004 ± 0.005) mm between both evaluations, respectively. The QAMaster results for spatial resolution did not differ significantly (p = 0.34) from the CatPhan® 504 based manual resolution assessment. Dose computations were fully consistent between QAMaster and manual calculations. Thus, QAMaster proved to be a comprehensive and functional software for performing an automated CT quality assurance routine. QAMaster will be open-source after its release.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomografia Computadorizada de Feixe Cônico Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomografia Computadorizada de Feixe Cônico Limite: Humans Idioma: En Revista: J Appl Clin Med Phys Assunto da revista: BIOFISICA Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Estados Unidos