3D Low-Cost Equipment for Automated Peritoneal Dialysis Therapy.
Rivero-Urzua, Samuel; Paredes-Rojas, Juan Carlos; Méndez-García, Sergio Rodrigo; Ortiz-Hernández, Fernando Eli; Oropeza-Osornio, Armando; Torres-SanMiguel, Christopher René.
Afiliação
  • Rivero-Urzua S; Instituto Politécnico Nacional, Escuela Superior de Ingeniería Mecánica y Eléctrica Unidad Zacatenco, Sección de Estudios de Posgrado e Investigación, Ciudad de Mexico 07738, Mexico.
  • Paredes-Rojas JC; Instituto Politécnico Nacional, Centro Mexicano para la Producción más Limpia, Acueducto de Guadalupe S/N, La laguna Ticomán, Ciudad de Mexico 07340, Mexico.
  • Méndez-García SR; Instituto Politécnico Nacional, Escuela Superior de Ingeniería Mecánica y Eléctrica Unidad Zacatenco, Sección de Estudios de Posgrado e Investigación, Ciudad de Mexico 07738, Mexico.
  • Ortiz-Hernández FE; Instituto Politécnico Nacional, Escuela Superior de Ingeniería Mecánica y Eléctrica, Unidad Culhuacán, Ciudad de Mexico 04260, Mexico.
  • Oropeza-Osornio A; Instituto Politécnico Nacional, Escuela Superior de Ingeniería Mecánica y Eléctrica, Unidad Ticomán, Ciudad de Mexico 07340, Mexico.
  • Torres-SanMiguel CR; Instituto Politécnico Nacional, Escuela Superior de Ingeniería Mecánica y Eléctrica Unidad Zacatenco, Sección de Estudios de Posgrado e Investigación, Ciudad de Mexico 07738, Mexico.
Healthcare (Basel) ; 10(3)2022 Mar 17.
Article em En | MEDLINE | ID: mdl-35327042
A breakthrough in peritoneal dialysis (PD) therapy occurred in 1977 with the development of continuous ambulatory peritoneal dialysis (CAPD). Its simplicity, low cost, and ease with which CAPD could be performed on patients at home contributed to the popularity of this procedure. However, there is a need for continuous improvement in building optimal systems for incident chronic kidney disease (CKD) patients. This research showed the design and construction of a simplified prototype of low-cost automated peritoneal dialysis (APD) equipment that meets international standards to automatically regulate infusion and fluid drainage in and out of a patient with low margins of error. Experimental tests allowed the adjustment of the RPM values concerning the flow rate provided. In addition, thanks to the pressure sensor, it was possible to observe a fluctuation ranging from 9 to 13 kPa, which is within the permissible average specified in the catalogs of medical instruments and equipment. Furthermore, a turbidity sensor was added to decrease the possibility of presenting peritonitis. The results showed absolute values of flow, angular velocity, and pressure that it could deliver for use in APD therapies. Finally, the construction of the APD equipment is presented generally, showing the electronic and mechanical components that constitute it.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Guideline / Health_economic_evaluation Idioma: En Revista: Healthcare (Basel) Ano de publicação: 2022 Tipo de documento: Article País de afiliação: México País de publicação: Suíça