Colorimetric and Reverse Fluorescence Dual-Signal Readout Immunochromatographic Assay for the Sensitive Determination of Sibutramine.
ACS Omega
; 9(6): 7075-7084, 2024 Feb 13.
Article
em En
| MEDLINE
| ID: mdl-38371773
ABSTRACT
Later flow immunochromatographic assay has been widely used in clinical, environmental, and other diagnostic applications owing to its high sensitivity and throughput. However, most immunoassays operate in the "turn-off" mode for detecting targets of low molecular weight. The signal intensity decreases as the analyte concentration increases, which poses a challenge for achieving ultrasensitive detection at low concentrations and is counterintuitive to new users. In this work, a fluorometric immunochromatographic assay (FICA) is developed to simultaneously read "turn-on" fluorescent and "turn-off" colorimetric signals, where ZnCdSe/ZnS quantum dots act as fluorescence donors and gold nanoparticles (AuNPs) act as quenchers. The fluorescent signal (excitation/emission wavelengths of 365/525 nm) is positively correlated with analytes' concentration. Taking sibutramine (SBT) as the analysis target, the visual limit of detection for SBT reached 3.9 ng/mL, and the limit of Quantitation was 5.0 ng/mg in spiked samples. The developed FICA achieves a high sensitivity in SBT detection, which is much lower than that of the colloidal gold-based immunochromatographic assay. This dual-function detection mode has great potential to be used as a rapid on-site semiquantitative method, providing an alternative mode for the determination of low levels of target analytes.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Idioma:
En
Revista:
ACS Omega
Ano de publicação:
2024
Tipo de documento:
Article
País de publicação:
Estados Unidos