Your browser doesn't support javascript.
loading
Endobronchial Closure for Peripheral Pulmonary Air Leakage.
Okado, Shoji; Nakamura, Shota; Sato, Keiyu; Ueno, Harushi; Ito, Takayasu; Sato, Kazuhide; Okachi, Shotaro; Uehara, Yohei; Matsumoto, Misa; Shimazaki, Natsumi; Miyagawa, Katsuya; Chen-Yoshikawa, Toyofumi Fengshi.
Afiliação
  • Okado S; Department of Thoracic Surgery, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Nakamura S; Department of Thoracic Surgery, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Sato K; Department of Thoracic Surgery, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Ueno H; Department of Thoracic Surgery, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Ito T; Department of Respiratory Medicine, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Sato K; Department of Respiratory Medicine, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Okachi S; Institute for Advanced Research (IAR), Nagoya University, Nagoya, Japan.
  • Uehara Y; Department of Respiratory Medicine, Nagoya University Graduate School of Medicine, Nagoya, Japan.
  • Matsumoto M; Department 5 for Research and Development, Research Institute, Nipro Corporation, Osaka, Japan.
  • Shimazaki N; Department 5 for Research and Development, Research Institute, Nipro Corporation, Osaka, Japan.
  • Miyagawa K; Department 5 for Research and Development, Research Institute, Nipro Corporation, Osaka, Japan.
  • Chen-Yoshikawa TF; Department 5 for Research and Development, Research Institute, Nipro Corporation, Osaka, Japan.
Respiration ; 103(10): 634-640, 2024.
Article em En | MEDLINE | ID: mdl-39079506
ABSTRACT

INTRODUCTION:

A minimally invasive alternative to surgery for treating pneumothorax has been developed, aiming to reduce risks while maintaining efficacy. This study conducted basic experiments using ex vivo and in vivo pig lung employing a super-thin catheter for treatment. This new device injects fibrin glue directly into the responsible lesion to close the air leak, which has two features thin design and double-lumen.

METHODS:

The experimental setup involved utilizing trachea and both lung specimens from pigs under positive pressure ventilation. To simulate pneumothorax, artificial fistulas were created on the lung surfaces. The super-thin catheter, guided through a bronchoscope near the fistula, was used to embolize the peripheral bronchus by injecting a fibrin preparation. Then, an air leak test was conducted afterward to assess the efficacy of the treatment. Additionally, a similar pneumothorax model was induced in alive pig under general anesthesia to evaluate its curability.

RESULTS:

In the extracted pig lungs, embolization was performed in 21 cases, resulting in the cessation of air leaks in 19 cases, corresponding to a 90.5% cure rate. Notably, no major adverse events occurred with the treatment devices. Similarly, in living pigs, pneumothorax was successfully treated, with no recurrence observed up to the seventh postoperative day.

CONCLUSION:

The novel treatment device utilizing a super-thin catheter offers a minimally invasive and highly curative option for pneumothorax. These promising results suggest the potential for further development and human clinical trials, which could revolutionize the treatment of pneumothorax, reducing risks and improving outcomes.