The RM Press fit cup™: an investigation in 182 hips at ten-year follow-up.
Portet, Adrien; Besnard, Marion; Ratsimbazafy, Carole; Berhouet, Julien; Samargandi, Ramy; Le Nail, Louis-Romée.
Afiliação
  • Portet A; Service de Chirurgie Orthopédique, Hôpital Trousseau, CHRU de Tours, Avenue de la République, Chambray-lès-Tours, Tours Cedex 9, 37044, France. Electronic address: adrien.portet@outlook.fr.
  • Besnard M; Centre Hospitalier Intercommunal Amboise-Château-Renault, Hôpital Robert-Debré, rue des Ursulines, BP 329, 37403 Amboise Cedex, France.
  • Ratsimbazafy C; Hôpital Saint-Antoine, Pharmacie, GHU AP-HP Sorbonne Université 184 rue du Faubourg-Saint-Antoine, 75012 Paris, France.
  • Berhouet J; Service de Chirurgie Orthopédique, Hôpital Trousseau, CHRU de Tours, Avenue de la République, Chambray-lès-Tours, Tours Cedex 9, 37044, France.
  • Samargandi R; Service de Chirurgie Orthopédique, Hôpital Trousseau, CHRU de Tours, Avenue de la République, Chambray-lès-Tours, Tours Cedex 9, 37044, France; Department of Orthopedic Surgery, Faculty of Medicine, University of Jeddah, Jeddah, Saudi Arabia.
  • Le Nail LR; Service de Chirurgie Orthopédique, Hôpital Trousseau, CHRU de Tours, Avenue de la République, Chambray-lès-Tours, Tours Cedex 9, 37044, France.
Orthop Traumatol Surg Res ; : 103988, 2024 Sep 06.
Article em En | MEDLINE | ID: mdl-39245266
ABSTRACT

INTRODUCTION:

For over 10 years, the RM Pressfit cup™ has been used in our department. This is a one-piece, elastic, cementless implant designed with standard polyethylene (PE), covered with a thin coating of titanium particles. To date, there is no French study evaluating this cup after more than 10 years. Therefore, we conducted a retrospective study in order to (1) evaluate the survival of the implant with a minimum follow-up of 10 years, (2) evaluate the functional scores at the last follow-up, (3) measure the wear of the PE, (4) identify radiological loosening, (5) search for risk factors for cup removal, (6) identify complications that required management in the operating theatre.

HYPOTHESIS:

The working hypothesis was that the survival of this implant was greater than 95% at 10 years' follow-up, in accordance with the criteria of the National Institute for Health and Care Excellence (NICE). MATERIALS AND

METHODS:

This was a retrospective monocentric study, including adult patients who underwent total hip arthroplasty (THA) with an RM Pressfit cup™ (28 mm friction size) for coxarthrosis (primary or secondary) or femoral head osteonecrosis. Exclusion criteria were a follow-up period of less than 10 years, the placement of an RM Pressfit cup™ as a secondary intention for a THA (n = 5) or following a trochanteric fracture (n = 1). In total, 163 patients (182 hips) with a median age (Q1-Q3) of 63 (56-68) years, and a sex ratio (M/F) of 1.7 were included. Functional scores were evaluated using the Harris and Oxford scores. Radiographs were analyzed in the immediate postoperative period and at the last follow-up.

RESULTS:

The median follow-up was 10.5 (10-11.5) years. Of the 182 included hips, 7 cups were removed, corresponding to a 10-year survival rate of 96.1% (95% CI [93.3; 96.9]). The median Harris and Oxford scores at 10.5 years were 95 (90-98) and 19 (17-23) points, respectively. The median PE wear rate was 0.058 (0.039-0.087) mm/year. Univariate analysis showed that male gender was associated with PE wear (OR = 3.6; 95% CI [1.3; 12.9] [p = 0.012]). Ten cups (6%) showed radiological instability with migration greater than 3 mm and/or variation in inclination greater than 8 °, and only 9 hips (6%) showed bone resorption. No preoperative or perioperative factors analyzed were associated with cup removal. Dislocation accounted for 71% (n = 5) of the causes of cup removal. Additionally, 6 hips experienced at least one dislocation episode requiring reduction by external maneuvers in the operating room, bringing the overall dislocation rate in the series to 6% (n = 11). Increased cup inclination was the only risk factor for prosthetic dislocation (OR = 1.2; 95% CI [1.09; 1.4] [p = 0.0003]). Overall complications requiring surgical intervention included 15 (8.3%) implanted cups (7 removed cups, 6 dislocation episodes requiring reduction by external maneuvers in the operating room, and 2 hips reoperated for washing and changing of mobile components due to early infection).

CONCLUSION:

The RM Pressfit cup™ gives good long-term clinical and radiological results with an overall survival of 96.1% and a low complication rate over 10 years. Over the last 2 years the RM Vitamys™ cup has been introduced allowing the use of 32 mm femoral head diameter for size 48 cups, to reduce the risk of dislocation. LEVEL OF EVIDENCE IV; retrospective cohort.