Your browser doesn't support javascript.
loading
A validated LC-MS/MS method for the simultaneous quantification of iothalamate and hippuran in serum and urine for non-radioactive kidney function assessment.
Mohamed, Abdulfataah A A; Walland, Peter; Stevens, Jasper; van Londen, Marco; Heerspink, Hiddo J L; Gansevoort, Ron T; van de Merbel, Nico C.
Afiliação
  • Mohamed AAA; Department of Clinical Pharmacy and Pharmacology, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713 GZ Groningen, The Netherlands; Department of Nephrology, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713 GZ Groningen, The Netherlands.
  • Walland P; ICON Bioanalytical Laboratories, Amerikaweg 18, 9407 TK Assen, The Netherlands.
  • Stevens J; Department of Clinical Pharmacy and Pharmacology, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713 GZ Groningen, The Netherlands.
  • van Londen M; Department of Nephrology, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713 GZ Groningen, The Netherlands.
  • Heerspink HJL; Department of Clinical Pharmacy and Pharmacology, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713 GZ Groningen, The Netherlands.
  • Gansevoort RT; Department of Nephrology, University of Groningen, University Medical Center Groningen, Hanzeplein 1, 9713 GZ Groningen, The Netherlands.
  • van de Merbel NC; ICON Bioanalytical Laboratories, Amerikaweg 18, 9407 TK Assen, The Netherlands; Department of Analytical Biochemistry, University of Groningen, Antonius Deusinglaan 1, 9713 AV Groningen, The Netherlands. Electronic address: n.c.van.de.merbel@rug.nl.
Article em En | MEDLINE | ID: mdl-39378524
ABSTRACT
A novel liquid chromatography-tandem mass spectrometry method is described for the quantitative determination of the kidney function markers iothalamate and hippuran in human serum and urine. It is based on protein precipitation with methanol followed by dilution of the supernatant for serum and simple dilution for urine. The polar analytes are chromatographically separated by a 6.5-min gradient on a low-ligand density reversed-phase column; detection is performed by electrospray ionization tandem mass spectrometry in the positive ion mode against stable-isotope labeled internal standards. The results of a thorough method validation show that iothalamate and hippuran can be simultaneously quantified in the concentration ranges 0.500-30.0 ng/mL and 10.0-5000 ng/mL for serum and urine, respectively, with values for CV and absolute bias not exceeding 10 %, and with sufficient stability in all relevant matrices and solvents. The method was successfully applied for the analysis of serum and urine samples of multiple individuals who received both iothalamate and hippuran.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci / J. chromatogr. B / Journal of chromatography. B (Print) Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Holanda País de publicação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci / J. chromatogr. B / Journal of chromatography. B (Print) Assunto da revista: ENGENHARIA BIOMEDICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Holanda País de publicação: Holanda