ABSTRACT
The working principle of F-50 helium compressor of Philips Ingenia 3.0T MRI Scanner was described.The causes and treatment of F-50 helium compressor were introduced.The importance of routine helium compressor maintenance for stable MRI operation was presented.References were provided for eliminating similar helium compressor faults.[Chinese Medical Equipment Journal,2023,44(10)114-117]