Your browser doesn't support javascript.
loading
Study on the effects of AI reconstruction algorithm based on different tube voltages on CT imaging quality and radiation dose of chest phantom / 中国医学装备
China Medical Equipment ; (12): 19-23, 2024.
Article em Zh | WPRIM | ID: wpr-1026517
Biblioteca responsável: WPRO
ABSTRACT
Objective:To investigate the effects of different tube voltages combined with artificial intelligence reconstruction algorithm(CI)on the computed tomography(CT)imaging quality and radiation dose of chest phantom on the basis of the CT scan for an adult male simulated chest phantom(PH-N1).Methods:A 512-slice CT scanner of ultrahigh-end was adopted to conduct scan,and the images were divided into 70 kV group,80 kV group,100 kV group and 120 kV group according to different tube voltage.For 4 groups of CT scan images with different tube voltages,the 10%,30%,50%,70%and 90%CI were adopted to reconstruct 1mm thin layer image.The CT volume dose index(CTDIvol)and the dose-length product(DLP)of the scans of 4 groups were compared.The CT values and standard deviation(SD)values of the aorta,abdominal wall fat and erector spine muscle were measured.Two senior diagnostic physicians with more than 5 years of work experience independently and double-blindly evaluated the image quality by using 5-point scale.A Kappa consistency test was conducted.One-way analysis of variance was adopted to compare the differences of CT values and SD values of the tissues of image targets.The Friedman rank-sum test was adopted to compare the differences of subjective image qualities among different groups.Results:The differences of CTDIvol and DLP among 4 groups with different tube voltages were significant(F=1855.617,3996.118,P<0.05),respectively.Under 70 kV tube voltage,there were no significant differences in CT values of the aorta,abdominal wall fat and erector spine muscle,which were reconstructed by using 10%,30%,50%,70%and 90%CI(P>0.05),while the differences of SD values among them were statistically significant(F=32.267,53.327,14.873,P<0.05),respectively.Under the different tube voltages of 4 groups,which were reconstructed by 90%CI,the CT values of aorta,abdominal wall fat and erector spine muscle gradually decreased with decreasing of tube voltage,the differences were significant(F=139.899,2563.93,219.231,P<0.05),respectively.The consistency of subjective scores between two diagnostic physicians was better for each group of images(Kappa=0.712~0.869).Conclusion:Compared with 80 kV,90 kV and 120 kV images,the reconstructed images with 90%CI algorithm under 70 kV tube voltage can significantly reduce the radiation dose,and the images have a favorable signal-to-noise ratio at the same time.
Palavras-chave
Texto completo: 1 Base de dados: WPRIM Idioma: Zh Revista: China Medical Equipment Ano de publicação: 2024 Tipo de documento: Article
Texto completo: 1 Base de dados: WPRIM Idioma: Zh Revista: China Medical Equipment Ano de publicação: 2024 Tipo de documento: Article