Ventilation impairment caused by upside-down connected unidirectional valves: A case report / 대한마취과학회지
Korean Journal of Anesthesiology
; : 736-739, 2008.
Artigo
em Coreano
| WPRIM (Pacífico Ocidental)
| ID: wpr-152767
Biblioteca responsável:
WPRO
ABSTRACT
We experienced a case of ventilation impairment resulted by the wrong-assembled unidirectional valves passed the automatic checkout of ventilator during general anesthesia. A 63-year-old man was scheduled for Ivor-Lewis operation due to esophageal cancer. After endotracheal intubation, ventilation impairment and abdominal distension were developed. No abnormal finding was found with bronchofiberoscopic examination. Suspecting anesthetic equipment failure, unidirectional valves were separated and they were proved to be connected upside-down.
Texto completo:
Disponível
Base de dados:
WPRIM (Pacífico Ocidental)
Assunto principal:
Ventilação
/
Neoplasias Esofágicas
/
Ventiladores Mecânicos
/
Falha de Equipamento
/
Intubação Intratraqueal
/
Anestesia Geral
Limite:
Humanos
Idioma:
Coreano
Revista:
Korean Journal of Anesthesiology
Ano de publicação:
2008
Tipo de documento:
Artigo