Your browser doesn't support javascript.
loading
In vitro pressure drop comparison between two mechanical valve prostheses
Yonsei Medical Journal ; : 72-76, 1994.
Artigo em Inglês | WPRIM (Pacífico Ocidental) | ID: wpr-171808
Biblioteca responsável: WPRO
ABSTRACT
An hemodynamic evaluation of two mechanical heart valves is presented. A tilting disc valve and a bileaflet valve were incorporated in a mock circulatory system which consists of a closed flow loop with a pneumatically driven flexible diaphragm to simulate the physiologic pulsatile flow. Comparisons between the valves were made on the aortic pressure, ventricular pressure, as well as mean pressure gradient at a systolic duration of 45% and a heart rate of 90 beats per minute. The results showed that the tilting disc valve has higher ventricular pressure and mean pressure gradient than that of the bileaflet valve. This indicates that the tilting disc valve has higher transvalvular flow resistance and energy loss than that of the bileaflet valve. From this study it is demonstrated that the mock circulatory system can be a very useful device to evaluate the prosthetic heart valves in vitro.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Pressão / Próteses Valvulares Cardíacas / Estudo Comparativo / Modelos Cardiovasculares / Modelos Estruturais Idioma: Inglês Revista: Yonsei Medical Journal Ano de publicação: 1994 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Pressão / Próteses Valvulares Cardíacas / Estudo Comparativo / Modelos Cardiovasculares / Modelos Estruturais Idioma: Inglês Revista: Yonsei Medical Journal Ano de publicação: 1994 Tipo de documento: Artigo
...