Your browser doesn't support javascript.
loading
Airway Obstruction by Displaced Cuff of Armored Tube: A case report / 대한마취과학회지
Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-192737
Biblioteca responsável: WPRO
ABSTRACT
A thirty nine year-old male patient was scheduled for emergent craniotomy and epidural hematoma evacuation under general anesthesia. All data of the preoperative check lists for emergent operation were within normal limits and vital signs of the patient were stable. During the induction of anesthesia an armored tube was inserted into the trachea for this operation. Approximately 110 minutes after the induction of anesthesia, signs of complete airway obstruction developed abruptly. At that time we deflated the cuff balloon to exchange the tube. Then, the signs of airway obstruction disappeared completely. Therefore the deflation of cuff balloon is recommended in order to exclude the obstruction by the displaced cuff.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Traqueia / Craniotomia / Obstrução das Vias Respiratórias / Sinais Vitais / Hematoma / Intubação / Anestesia / Anestesia Geral Limite: Humanos / Masculino Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 1996 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Traqueia / Craniotomia / Obstrução das Vias Respiratórias / Sinais Vitais / Hematoma / Intubação / Anestesia / Anestesia Geral Limite: Humanos / Masculino Idioma: Coreano Revista: Korean Journal of Anesthesiology Ano de publicação: 1996 Tipo de documento: Artigo
...