Your browser doesn't support javascript.
loading
Usefulness of modified ambu(R) in patients who need artificial ventilation / 소아과
Korean Journal of Pediatrics ; : 1194-1201, 2006.
Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-19358
Biblioteca responsável: WPRO
ABSTRACT

PURPOSE:

The comatose mentality can be catastrophic, especially if the condition is severe or the duration is prolonged. Therefore, delayed diagnosis can result in a poor outcome or death. The best radiologic modality to differentiate from cerebral lesions in patients suffering from cerebral diseases is magnetic resonance imaging (MRI) rather than computed tomography (CT). Special apparatuses with metal materials such as ventilators, and cardiac pacemakers belonging to patients cannot be located in the magnetic field. We aimed to exhibit the possibility of examining MRI, maintaining ventilation at a relative long distance by means of modified Ambu(R).

METHODS:

Self-inflating bags as a sort of a manual ventilator, connected with relatively long extension tubes instead of mechanical ventilators, were adopted to obtain MRI. PVC (polyvinyl chloride) extension tubes had different lengths and diameters. Lengths were 1, 2, and 3 cm and diameters were 15, and 25 mm. The work of breathing and expiratory changes of expiratory tidal volume (TVe), minute volume of expiration (MVe), peak inspiratory pressure (PIP) were measured by use of the mechanical ventilator, Servoi(R), as the alteration of TVi (inspiratory tidal volume), extension tube lengths and diameters with other values fixed.

RESULTS:

Measured TVe and MVe by ventilator were the same values with control at every TVi, regardless of extension tube lengths and diameters, but PIP were increased with the rise of TVi, tube lengths, with decline of tube diameters, these were statistically significant.

CONCLUSION:

MRI examination can be carried out with a self-inflating bag connected with an extension tube at a long distance in patients who need artificial ventilation.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Ventilação / Imageamento por Ressonância Magnética / Ventiladores Mecânicos / Volume de Ventilação Pulmonar / Trabalho Respiratório / Coma / Diagnóstico Tardio / Campos Magnéticos Tipo de estudo: Estudo diagnóstico / Guia de prática clínica Limite: Humanos Idioma: Coreano Revista: Korean Journal of Pediatrics Ano de publicação: 2006 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Ventilação / Imageamento por Ressonância Magnética / Ventiladores Mecânicos / Volume de Ventilação Pulmonar / Trabalho Respiratório / Coma / Diagnóstico Tardio / Campos Magnéticos Tipo de estudo: Estudo diagnóstico / Guia de prática clínica Limite: Humanos Idioma: Coreano Revista: Korean Journal of Pediatrics Ano de publicação: 2006 Tipo de documento: Artigo
...