Your browser doesn't support javascript.
loading
Pacemaker malfunction related to insulation break according to position change in a patient with complete AV block / 대한내과학회지
Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-203637
Biblioteca responsável: WPRO
ABSTRACT
A 53 year-old woman presented with intermittent dizziness and palpitation. She had received VVI type pacemaker due to complete AV block in 1990, and exchanged by VVIR type pacemaker thirteen years later. 1 year later, she suffered intermittent dizziness and palpitation in erect position, not in supine position. Intermittent pacing failure and sensing failure was observed in pacemaker test, especially in erect position, not in supine position. Pacing threshold was increased in erect position, but lead impedance was not changed. Insulation break was observed in bipolar lead by fluoroscope. Initially we tried a new lead implantation by cephalic access to prevent lead related complication, but failed. So, we implanted a new bipolar lead by subclavian access. Finally, she was treated by a new bipolar lead implantation.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Estimulação Cardíaca Artificial / Decúbito Dorsal / Impedância Elétrica / Tontura / Bloqueio Atrioventricular Limite: Feminino / Humanos Idioma: Coreano Revista: Korean Journal of Medicine Ano de publicação: 2006 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Estimulação Cardíaca Artificial / Decúbito Dorsal / Impedância Elétrica / Tontura / Bloqueio Atrioventricular Limite: Feminino / Humanos Idioma: Coreano Revista: Korean Journal of Medicine Ano de publicação: 2006 Tipo de documento: Artigo
...