Dosimetric comparison of treatment plans of pancreatic carcinoma treated with body gamma knife and tomotherapy / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
; (6): 232-234, 2013.
Artigo
em Chinês
| WPRIM (Pacífico Ocidental)
| ID: wpr-264225
Biblioteca responsável:
WPRO
ABSTRACT
Fourteen patients with pancreatic carcinoma were selected. Two treatment plans were designed for each patient, including gamma knife and Tomotherapy. The dose characteristics were evaluated by DVH and were compared. The results showed that the gamma knife plan had the higher maximal and mean target dose than Tomotherapy. Body gamma knife can increase the target dose significantly, and decrease the OAR dose. Tomotherapy had excellent dose-target conformality, and it can control doses of duodenum and stomach easily, but it had larger low dose region.
Texto completo:
Disponível
Base de dados:
WPRIM (Pacífico Ocidental)
Assunto principal:
Neoplasias Pancreáticas
/
Radioterapia
/
Dosagem Radioterapêutica
/
Planejamento da Radioterapia Assistida por Computador
/
Radiocirurgia
/
Métodos
Limite:
Humanos
Idioma:
Chinês
Revista:
Chinese Journal of Medical Instrumentation
Ano de publicação:
2013
Tipo de documento:
Artigo