Digital fluoroscopy imaging system with dual energy X-ray image decomposition / 生物医学工程学杂志
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-291130
Biblioteca responsável: WPRO
ABSTRACT
Dual-kilovolt technique was realized in dual energy X-ray imaging system with computer controlled KV adjustment, filter switching and image acquirement. The dual energy images can be obtained in 2-3 seconds to avoid moving artifact. Low energy photons in high energy X-ray beam can be attenuated effectively by a 2 mm Cu filter to separate the spectra of the high and low energy X-rays. A simplified dual energy decomposition method was used to obtain the soft tissue image. An example of chest dual energy radiographs is presented to illustrate the performance of the system.
Assuntos
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Processamento de Sinais Assistido por Computador / Radiografia Torácica / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Métodos Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2004 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Processamento de Sinais Assistido por Computador / Radiografia Torácica / Imagem Radiográfica a Partir de Emissão de Duplo Fóton / Métodos Limite: Humanos Idioma: Chinês Revista: Journal of Biomedical Engineering Ano de publicação: 2004 Tipo de documento: Artigo
...